15 Transport and Storage
Instruction Manual Mesotest II (G/62800/0715/en)
75 / 86
15 Transport and Storage
The Mesotest II, must be properly dismantled and packed before being
transported or stored.
15.1 Transport and Storage Information
Storage
This device does not keep up with the temperature conditions for storage
according to ISO 15004-1.
Do not store the Mesotest II under conditions over 50 °C and under
-10 °C.
In accordance with DIN EN 60601-1, the storage conditions are:
Transport
In accordance with DIN EN 60601-1, the transport conditions are:
After storage and transport
Wait approx. 3-4 hours after transport before operating the
Mesotest II for the first time. In the event of extreme temperature
changes from cold areas to warm rooms, the optical components can
become fogged.
Ambient temperature range
-10°C to +55°C
Relative humidity, including condensation
10% to 95%
Air pressure range
700 hPa to1060 hPa
Ambient temperature range
-40°C to +70°C
Relative humidity, including condensation
10% to 95%
Air pressure range
500 hPa to 1060 hPa