41
SET END PRESSURE ALARM
information includes (Fig. 33):
> Graphic END
> Set Point numeric value, flashing.
> Symbol PSI (or M)
• Pressing and releasing the S button momentarily (< 2
seconds) will increase the Set Point from 300 to 1500 PSI
(20 to 105 BAR) in 100 PSI (5 BAR) increments.
• Depressing the S button will scroll through the Set Points at a
rate of 4 Set Points per second until it is released.
• Pressing and releasing the A button momentarily (< 2
seconds) will save the setting and/or advance to the SET
PO2 ALARM screen with the Set Point flashing.
• Pressing and releasing the A button repeatedly (< 2 seconds
each) will step through the other SET A screens.
• Depressing the A and S buttons simultaneously for 2 seconds
will save the setting and revert to the SET A screen.
• Depressing the M button for 2 seconds or if no button is
pressed for a period of 2 minutes operation will revert to the
NORM or GAUG SURF MAIN screen.
Fig. 33 - Set End Alarm
The END PRESSURE
Alarm will activate
when Pressure in the
Tank (TMT 1, 2 , or
3) being used at the
time decreases to the
Alarm Set Point.