7.2.
Finding the correct wiring diagram for
your sensor
To find out the appropriate wiring diagram for calibration and general use, the user may refer
to the table on
Chapter 9 “PINOUT & WIRING DIAGRAM”
. This selection table also includes
reference to the recommended wiring diagrams via tag markers
(#WD1
to
#WD8
).
7.3.
How to configure and calibrate your
sensor
The NEP-5000 offers easy configuration methodology using single wire (one-wire)
communication to the configuration software. Using this configuration, the user only requires to
connect 3 wires during the time of the calibration process.
To facilitate trouble free communication with the PC, Observator Instruments provides a module
called “Calibration Module” that converts one wire calibration data to USB and generates
12VDC from available USB 5V.
Note:
The calibration module that plugs into the USB port must be able to provide more than
200mA at 5V. If not, the user may provide external power using 12VDC, 500mA plug-pack.