NEP-5000 MANUAL
V2019-08-29
Observator Instruments
Copyright 2019
©
Page 97
•
Press: “Set”, and “Get live data”, but the reading is still un-desirable, redo all previous steps
until the reading is close enough to the high top-range NTU value.
Figure 11.4.8.F: Set the high-range top calibration value
•
Press “Save calibration” to commit your new calibration data to the permanent memory inside
the sensor.
Figure 11.4.8.G: Set the high-range top calibration value
Note
: It is recommended that you export the newly calibrated data to a file for future use.