background image

15

14

Notes

When entering calibration mode, the meter will display the factory default value. If
the meter was previously calibrated, the display may “jump” to the factory default
value when switching from measurement to calibration mode. 

To exit from Conductivity Calibration mode without confirming calibration,

DO NOT

press ENTER in step 6. Press CAL/MEAS instead. This will retain the meter’s old calibra-
tion data in the measuring range of the calibration.

You can offset the conductivity reading up to ±20% from the default setting. If your
measured value differs by more than ±20%, clean or replace probe as needed. See
page 47 for a wide selection of high-quality calibration standards.

To calibrate conductivity:

1. 

If necessary, press the MODE key to
select conductivity mode

.

See figure

2. 

Rinse the probe thoroughly with 
de-ionized water or a rinse solution,
then rinse with a small amount of 
calibration standard.

3. 

Dip the probe into the calibration 
standard. 

Immerse the probe tip beyond

the upper steel band. Stir the probe gently
to create a homogeneous sample. 

See figure

4. 

Wait for the measured conductivity
value to stabilize.

If the READY indica-

tor has been activated (set up program
P7.1—see page 34), the READY annunci-
ator lights when the reading is stable.

5. 

Press CAL/MEAS

to enter conductivity

calibration mode. The CAL indicator 
will appear above the display. 

See figure

6. 

Press the 

or 

to change the value on

the primary display to match the value of
the calibration standard. 

See figure

7. 

Press ENTER

to confirm the calibration

value. The meter returns to the MEAS
(measurement) mode. 

8. 

Repeat steps 1-7 for other ranges.

CAL

ATC

READY

 C

µ

S

C

ATC

READY

CAL

 C

µ

S

D

D

ATC

READY

MEAS

 C

µ

S

A

A

B

B

C

Immerse
probe beyond
upper steel
band

1. 

If necessary, press the MODE key to
select TDS mode

.

See figure

2. 

Rinse the probe with de-ionized water or
a rinse solution,

then rinse with a small

amount of calibration standard.

3. 

Dip the probe into the calibration 
standard. 

Immerse the probe tip beyond

the upper steel band. Stir the probe gently
to create a homogeneous sample. 

See figure

4. 

Wait for the measured TDS value to
stabilize.

If the READY indicator has

been activated (set up program P7.1—
see page 34), the READY annunciator
lights when the reading is stable.

5. 

Press CAL/MEAS

to enter TDS calibra-

tion mode. The CAL indicator will appear
above the display. 

See figure

6. 

Press the 

or 

keys

to change the

value on the primary display to match the
value of the calibration standard. 

See figure

7. 

Press ENTER

to confirm the calibration

value. The meter returns to the MEAS
(measurement) mode. 

8. 

Repeat steps 1-7 for other ranges.

ATC

READY

CAL

 C

ppm

C

ATC

READY

CAL

 C

ppm

D

D

ATC

READY

MEAS

 C

ppm

A

A

B

B

C

Immerse
probe beyond
upper steel
band

Calibrating for TDS directly

The factory default setting for TDS conversion factor is 0.5. If your solution has a dif-
ferent TDS factor, you can improve calibration accuracy by setting the TDS factor
prior to calibration. See SetUp Program P7.4 on page 37 for directions. See
Appendixes 1 and 2 on pages 48-49 to determine your exact TDS factor. 

Notes

When entering calibration mode, the meter will display the factory default value. If
the meter was previously calibrated, the display may “jump” to the factory default
value when switching from measurement to calibration mode. 

To exit from TDS Calibration mode without confirming calibration,

DO NOT press

ENTER 

in step 6. Press CAL/MEAS instead. This will retain the meter’s old calibration

data in the measuring range of the calibration.

You can offset the TDS reading up to ±20% from the default setting. If your meas-
ured value differs by more than ±20%, clean or replace probe as needed.

Содержание 35631-00

Страница 1: ...me period please return freight prepaid and correction will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products w...

Страница 2: ...ady indicator auto endpoint function 28 P3 2 Selecting number of pH calibration points 29 P3 3 Selecting C or F 30 6 5 Program 4 0 Resetting to factory default settings pH 31 6 6 Program 5 0 Viewing p...

Страница 3: ...splay also shows error messages keypad functions and program functions mS ATC READY MEAS CAL HOLD ERR F SETUP ON OFF C S K pH mV pH ppt ppm 2 1 Primary display Secondary display 3 2 21 20 14 17 16 15...

Страница 4: ...ions In advanced set up mode Press CAL MEAS to return to main menu from sub menus Press CAL MEAS again to return to measurement mode from main menu ENTER Press to confirm values in Calibration mode an...

Страница 5: ...To connect the pH electrode 1 Slide the BNC connector of the probe over the BNC connector socket on the meter Make sure the slots of the connector are in line with the posts of the socket Rotate and...

Страница 6: ...eplaced on a point by point basis in pH and on a range by range basis in conductivity or TDS For example pH if you previously calibrated your meter at pH 4 01 7 00 and 10 01 and you recalibrate at pH...

Страница 7: ...each of the ranges you are measuring All new calibration data will over ride existing stored calibration data for each measuring range you calibrate If you are measuring in ranges near to or greater t...

Страница 8: ...and 1 If necessary press the MODE key to select TDS mode See figure 2 Rinse the probe with de ionized water or a rinse solution then rinse with a small amount of calibration standard 3 Dip the probe i...

Страница 9: ...nal reading To exit this program without confirming the temperature calibration value DO NOT press ENTER in step 7 Press CAL MEAS instead 5 Measurement To take readings 1 Rinse the pH electrode with d...

Страница 10: ...y to toggle between conductivity and pH readings Taking Conductivity or TDS Measurements ATC READY MEAS C S A A 5 2 B B Immerse probe beyond upper steel band Notes You can use the conductivity TDS man...

Страница 11: ...h time you turn the meter off 5 4 From measurement mode 1 Press the Mode key to select conductivity measurement mode 2 Press Setup key to enter Set Up mode 3 Press the and keys to scroll through subgr...

Страница 12: ...E C CAL F F The meter will now compensate readings for the manually set temperature Notes To exit this program without confirming the manual temperature compensation value DO NOT press ENTER in step...

Страница 13: ...onductivity TDS calibration data R2 R2 0 0 199 9 S 10 0 99 9 ppm P5 3 View conductivity TDS calibration data R3 R3 0 1999 S 100 999 ppm P5 4 View conductivity TDS calibration data R4 R4 0 00 19 99 mS...

Страница 14: ...electrode parameters for diagnostic purposes It lets you view P2 1 Electrode offset P2 2 Electrode slope SETUP pH A A SETUP mV pH B B 6 3 SETUP pH C C From measurement mode 1 Press the Mode key to se...

Страница 15: ...See figure 5 Press the and keys to select the configuration you require OFF switches the READY indicator off ON switches the READY indicator on ON and HOLD together switches the auto endpoint feature...

Страница 16: ...the ENTER key See figure 5 Press the and keys to toggle between NO and YES NO retains current settings YES resets to factory default settings 6 Press the ENTER key to confirm selection and to return...

Страница 17: ...to your calibration 6 7 mS SETUP S ppt ppm A A SETUP K B B From measurement mode 1 Press the Mode key to select conductivity or TDS measurement mode 2 Press the Set up key to enter Set Up mode See fig...

Страница 18: ...eter 7 1 See figure 5 Press the and keys to select the configuration you require ON switches the READY indicator on OFF switches the READY indicator off ON and HOLD together switches the auto endpoint...

Страница 19: ...solution See Appendix 2 on page 49 for information on how to calculate the TDS factor 7 Press the and keys to select your calculated TDS conversion factor 8 Press the ENTER key to confirm selection a...

Страница 20: ...e temperature coefficient is the amount of change in conductivity per degree of temperature it is expressed in percent per C or F Entering the exact temperature coefficient of your solution lets you a...

Страница 21: ...ode into the heated solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposits prepare a 1 pepsin solution in 0 1 M of HCl Set the electrode in the soluti...

Страница 22: ...nstead of performing this reactivation procedure 1 Dip or stir the pH electrode in alcohol for 5 minutes 2 Leave the electrode in tap water for 15 minutes 3 Dip and stir the electrode in concentrate a...

Страница 23: ...al LCD Operating temperature 0 to 50 C 32 to 122 F Power four 1 5 V AAA batteries included Battery life 200 hours continuous use Probe connectors pH mV BNC connector Conductivity TDS Temperature notch...

Страница 24: ...00654 00 pH 4 01 calibration buffer 1 pint WD 00654 04 pH 7 01 calibration buffer 1 pint WD 00654 08 pH 10 01 calibration buffer 1 pint WD 00654 12 pH 12 45 calibration buffer 1 pint Model No Solution...

Страница 25: ...in the above formula The result should be the TDS value 12 Appendix 1 Conductivity to TDS Conversion Factors 1 Factor the conductivity to ppm TDS conversion factor Multiply conductivity by this factor...

Страница 26: ...the meter readings 15 Appendix 4 Meter factory default settings Resetting the meter to factory default settings clears all calibration data and most other setup functions you might have changed The fo...

Отзывы: