background image

13

12

4. 

Press CAL/MEAS

to enter pH calibration

mode. The CAL indicator lights. The 
primary display will show the measured
reading while the smaller secondary 
display will indicate the pH standard
buffer solution. 

See figure 

5. 

Wait for the measured pH value to 
stabilize.

If the READY indicator has

been activated (set up program P3.1—
see page 28), the READY annunciator
lights when the reading is stable.

6. 

Press ENTER to confirm calibration

The meter is now calibrated to the 
current buffer. The lower display scrolls
through the remaining buffer options.

• If you are performing multipoint 

calibration, go to step 7. 

• If you are performing one-point 

calibration, go to step 9.

7. 

Rinse the electrode with de-ionized
water or a rinse solution, and place it
in the next pH buffer. 

8. 

Follow steps 5 to 8 for additional 
calibration points. 

9. 

When calibration is complete, 
press CAL/MEAS to return to pH 
measurement mode

.

See figure 

Notes

To exit from pH Calibration mode without confirming calibration,

DO NOT press

ENTER 

in step 6. Press CAL/MEAS instead.

If the selected buffer value is not within ±1.0 pH from the measured pH value: the
electrode and buffer icon blink and the ERR annunciator appears in the lower left
corner of the display.

To limit the number of pH buffer values available during calibration, see Set-up 
program P3.2 on page 29.

ATC

READY

MEAS

 C

pH

C

C

READY

CAL

pH

pH

B

B

Conductivity and TDS calibration

Calibrate up to 5-point conductivity or 5 point TDS calibration at one point per range:  

Range

Conductivity:

TDS: 

R1

0.00-19.99 µS

0.00-9.99 ppm

R2

0.0-199.9 µS

10.0-99.9 ppm

R3

0-1999 µS

100-999 ppm

R4

0.00-19.99 mS

1.00-9.99 ppt

R5

0.0-199.9 mS

10.0-199.9 ppt

If you are measuring values in more than one range, make sure to calibrate each of 
the ranges you are measuring. All new calibration data will over-ride existing stored
calibration data for each measuring range you calibrate.

• If you are measuring in ranges near to or greater than 20 mS (10 ppt), or near to 

or lower than 100 µS (50 ppm), calibrate the meter at least once a week to get 
specified ±1% F.S. accuracy. 

• If you are measuring in the mid ranges and you washed the probe in deionized

water and stored it dry, calibrate the meter at least once a month. 

• If you take measurements at extreme temperatures, calibrate the meter at least

once a week. 

Preparing for conductivity/TDS calibration

For best results, select a standard value close to the sample value you are 
measuring. Alternatively, use a calibration solution value that is approximately 

2

3

the full scale value of the measurement range you plan to use. For example, 

in the 0 to 1999 µS conductivity range, use a 1413 µS solution for calibration.

See the table below for recommended calibration solution ranges:

4.3

Temperature coefficient:

These meters are factory set to a temperature coefficient 

of 2.1% per °C. For most applications this will provide good results. See 
Program P8.1 on page 38 to set the temperature coefficient to a different value. 
See Appendix 31, “Calculating Temperature Coefficients” on page 50 to determine 
the appropriate temperature coefficient for your solution.

Normalization temperature:

The factory default value for normalization tempera-

ture is 25°C. If you need to normalize to a value other than 25°C, see Program P8.2
on page 39.

CONTINUED ON NEXT PAGE

Conductivity

Recommended

TDS

Recommended 

Range 

Cal. Solution Range

Range

Cal. Solution Range

0.00-19.99 µS

6.00 to 17.00 µS

0.00-9.99 ppm

3.00 to 8.50 ppm

0.0-199.9 µS

60.0 to 170.0 µS

10.0-99.9 ppm

30.0 to 85.0 ppm

0-1999 µS

600 to 1700 µS

100-999 ppm

300 to 850 ppm

0.00-19.99 mS

6.00 to 17.00 mS

1.00-9.99 ppt

3.00 to 8.50 ppt

0.0-199.9 mS

60.0 to 170.0 mS

10.0-199.9 ppt

30.0 to 170 ppt

Содержание 35631-00

Страница 1: ...me period please return freight prepaid and correction will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products w...

Страница 2: ...ady indicator auto endpoint function 28 P3 2 Selecting number of pH calibration points 29 P3 3 Selecting C or F 30 6 5 Program 4 0 Resetting to factory default settings pH 31 6 6 Program 5 0 Viewing p...

Страница 3: ...splay also shows error messages keypad functions and program functions mS ATC READY MEAS CAL HOLD ERR F SETUP ON OFF C S K pH mV pH ppt ppm 2 1 Primary display Secondary display 3 2 21 20 14 17 16 15...

Страница 4: ...ions In advanced set up mode Press CAL MEAS to return to main menu from sub menus Press CAL MEAS again to return to measurement mode from main menu ENTER Press to confirm values in Calibration mode an...

Страница 5: ...To connect the pH electrode 1 Slide the BNC connector of the probe over the BNC connector socket on the meter Make sure the slots of the connector are in line with the posts of the socket Rotate and...

Страница 6: ...eplaced on a point by point basis in pH and on a range by range basis in conductivity or TDS For example pH if you previously calibrated your meter at pH 4 01 7 00 and 10 01 and you recalibrate at pH...

Страница 7: ...each of the ranges you are measuring All new calibration data will over ride existing stored calibration data for each measuring range you calibrate If you are measuring in ranges near to or greater t...

Страница 8: ...and 1 If necessary press the MODE key to select TDS mode See figure 2 Rinse the probe with de ionized water or a rinse solution then rinse with a small amount of calibration standard 3 Dip the probe i...

Страница 9: ...nal reading To exit this program without confirming the temperature calibration value DO NOT press ENTER in step 7 Press CAL MEAS instead 5 Measurement To take readings 1 Rinse the pH electrode with d...

Страница 10: ...y to toggle between conductivity and pH readings Taking Conductivity or TDS Measurements ATC READY MEAS C S A A 5 2 B B Immerse probe beyond upper steel band Notes You can use the conductivity TDS man...

Страница 11: ...h time you turn the meter off 5 4 From measurement mode 1 Press the Mode key to select conductivity measurement mode 2 Press Setup key to enter Set Up mode 3 Press the and keys to scroll through subgr...

Страница 12: ...E C CAL F F The meter will now compensate readings for the manually set temperature Notes To exit this program without confirming the manual temperature compensation value DO NOT press ENTER in step...

Страница 13: ...onductivity TDS calibration data R2 R2 0 0 199 9 S 10 0 99 9 ppm P5 3 View conductivity TDS calibration data R3 R3 0 1999 S 100 999 ppm P5 4 View conductivity TDS calibration data R4 R4 0 00 19 99 mS...

Страница 14: ...electrode parameters for diagnostic purposes It lets you view P2 1 Electrode offset P2 2 Electrode slope SETUP pH A A SETUP mV pH B B 6 3 SETUP pH C C From measurement mode 1 Press the Mode key to se...

Страница 15: ...See figure 5 Press the and keys to select the configuration you require OFF switches the READY indicator off ON switches the READY indicator on ON and HOLD together switches the auto endpoint feature...

Страница 16: ...the ENTER key See figure 5 Press the and keys to toggle between NO and YES NO retains current settings YES resets to factory default settings 6 Press the ENTER key to confirm selection and to return...

Страница 17: ...to your calibration 6 7 mS SETUP S ppt ppm A A SETUP K B B From measurement mode 1 Press the Mode key to select conductivity or TDS measurement mode 2 Press the Set up key to enter Set Up mode See fig...

Страница 18: ...eter 7 1 See figure 5 Press the and keys to select the configuration you require ON switches the READY indicator on OFF switches the READY indicator off ON and HOLD together switches the auto endpoint...

Страница 19: ...solution See Appendix 2 on page 49 for information on how to calculate the TDS factor 7 Press the and keys to select your calculated TDS conversion factor 8 Press the ENTER key to confirm selection a...

Страница 20: ...e temperature coefficient is the amount of change in conductivity per degree of temperature it is expressed in percent per C or F Entering the exact temperature coefficient of your solution lets you a...

Страница 21: ...ode into the heated solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposits prepare a 1 pepsin solution in 0 1 M of HCl Set the electrode in the soluti...

Страница 22: ...nstead of performing this reactivation procedure 1 Dip or stir the pH electrode in alcohol for 5 minutes 2 Leave the electrode in tap water for 15 minutes 3 Dip and stir the electrode in concentrate a...

Страница 23: ...al LCD Operating temperature 0 to 50 C 32 to 122 F Power four 1 5 V AAA batteries included Battery life 200 hours continuous use Probe connectors pH mV BNC connector Conductivity TDS Temperature notch...

Страница 24: ...00654 00 pH 4 01 calibration buffer 1 pint WD 00654 04 pH 7 01 calibration buffer 1 pint WD 00654 08 pH 10 01 calibration buffer 1 pint WD 00654 12 pH 12 45 calibration buffer 1 pint Model No Solution...

Страница 25: ...in the above formula The result should be the TDS value 12 Appendix 1 Conductivity to TDS Conversion Factors 1 Factor the conductivity to ppm TDS conversion factor Multiply conductivity by this factor...

Страница 26: ...the meter readings 15 Appendix 4 Meter factory default settings Resetting the meter to factory default settings clears all calibration data and most other setup functions you might have changed The fo...

Отзывы: