25
24
Advanced set-up mode detailed overview
P1.0: Viewing pH calibration data
P1.1 View previous pH calibration data (pH 1.68)
P1.2 View previous pH calibration data (pH 4.01)
P1.3 View previous pH calibration data (pH 7.00)
P1.4 View previous pH calibration data (pH 10.01)
P1.5 View previous pH calibration data (pH 12.45)
P2.0: Viewing pH electrode data
P2.1 View electrode offset
P2.2 View electrode slope
6.1
P3.0: pH configuration
P3.1 Ready indicator on or off /
auto endpoint on or off
P3.2 Select number of pH calibration points
P3.3 Select °F or °C
P4.0: Reset to factory default (pH)
Reset pH data to factory default settings
Instructions on page 26
Press the SETUP key to enter Set up mode. Press the
▲
and
▼
keys to scroll through sub groups.
CAL
SETUP
pH
SETUP
pH
SETUP
pH
MEAS
SETUP
pH
P9.0: Reset to factory default (Con./TDS)
Reset conductivity data to factory default settings
mS
MEAS
SETUP
µ
S
ppt
ppm
Sub groups available from pH measurement mode
P5.0: Viewing Con./TDS calibration data
P5.1 View conductivity/TDS calibration data R1
R1 = 0.00-19.99 µS / 0.00-9.99 ppm
P5.2 View conductivity/TDS calibration data R2
R2 = 0.0-199.9 µS / 10.0-99.9 ppm
P5.3 View conductivity/TDS calibration data R3
R3 = 0-1999 µS / 100-999 ppm
P5.4 View conductivity/TDS calibration data R4
R4 = 0.00-19.99 mS / 1.00-9.99 ppt
P5.5 View conductivity/TDS calibration data R5
R5 = 0.0-199.9 mS / 10.0-199.9 ppt
P6.0: Viewing Con./TDS probe data
P6.1 View effective cell constant R1
R1 = 0.00-19.99 µS / 0.00-9.99 ppm
P6.2 View effective cell constant R2
R2 = 0.0-199.9 µS / 10.0-99.9 ppm
P6.3 View effective cell constant R3
R3 = 0-1999 µS / 100-999 ppm
P6.4 View effective cell constant R4
R4 = 0.00-19.99 mS / 1.00-9.99 ppt
P6.5 View effective cell constant R5
R5 = 0.0-199.9 mS / 10.0-199.9 ppt
P7.0: Con./TDS configuration
P7.1 Ready indicator on or off /
auto endpoint on or off
P7.2 Select °F or °C
P7.3 Select ATC/Manual Temperature Compensation
P7.4 Set TDS factor
P8.0: Temperature parameters
P8.1 Set temperature coefficient
P8.2 Set normalization temperature
Instructions on page 32
mS
SETUP
µ
S
ppt
ppm
mS
SETUP
µ
S
ppt
ppm
mS
SETUP
µ
S
ppt
ppm
mS
F
SETUP
C
µ
S
ppt
ppm
Sub groups available from conductivity/TDS measurement mode
Instructions on page 27
Instructions on page 33
Instructions on pages 34-37
Instructions on pages 38-39
Instructions on pages 28-30
Instructions on page 31