background image

33

32

From measurement mode: 
1. 

Press the Mode key 

to select 

conductivity or TDS measurement
mode.

1. 

Press the Set up key 

to enter Set Up

mode.

2.

Press the 

and 

keys

to scroll

through subgroups until you view
parameter P6.0. 

See figure

3. 

Press the ENTER key repeatedly
to view the effective cell constant for
each range. 

Range

Conductivity:

TDS: 

P6.1 (R1) 0.00-19.99 µS

0.00-9.99 ppm

P6.2 (R2) 0.0-199.9 µS

10.0-99.9 ppm

P6.3 (R3) 0-1999 µS

100-999 ppm

P6.4 (R4) 0.00-19.99 mS

1.00-9.99 ppt

P6.5 (R5) 0.0-199.9 mS

10.0-199.9 ppt

See figure     

4.

When you have scrolled through all 
probe data, you will automatically
return to the subgroup menu. Press 
the CAL/MEAS key to return to 
measurement mode.

Notes

Cell constants will degrade with time and
usage. You can use this feature to alert you
to when your probe needs cleaning or to
when you need to replace your probe.

P6.0: Viewing conductivity/TDS probe data

Program 3 has five "view only" options that let you check your conductivity/TDS
probe’s parameters for diagnostic purposes. These options show you the effective cell
constant for each range. The cell constant is adjusted according to your calibration.

6.7

mS

SETUP

µ

S

ppt

ppm

A

A

SETUP

K=

B

B

From measurement mode: 

1. 

Press the Mode key 

to select 

conductivity or TDS measurement
mode.

2. 

Press the Set up key 

to enter Set Up

mode.

See figure

3.

Press the 

and 

keys

to scroll

through subgroups until you view
parameter P5.0. 

4. 

Press the ENTER key repeatedly

to

view previous calibration data.

Range

Conductivity:

TDS: 

P5.1 (R1) 0.00-19.99 µS

0.00-9.99 ppm

P5.2 (R2) 0.0-199.9 µS

10.0-99.9 ppm

P5.3 (R3) 0-1999 µS

100-999 ppm

P5.4 (R4) 0.00-19.99 mS

1.00-9.99 ppt

P5.5 (R5) 0.0-199.9 mS

10.0-199.9 ppt

See figure

5.

When you have scrolled through all 
calibration data, you will automatically
return to the subgroup menu. Press the
CAL/MEAS key to return to measure-
ment mode.

Notes

If there is no previous calibration data at a
particular point, the primary display will
show “– – –”.

See figure

P5.0: Viewing previous conductivity/TDS calibration data

This mode lets you recall previous conductivity or TDS calibration data, which lets
you know at which points this meter was previously calibrated. This is a “view only”
mode. 

TDS values are proportional to conductivity values. Note that if you calibrate a TDS
value in an equivalent conductivity range, the TDS value will replace the previous
conductivity value, and vice versa. 

6.6

mS

SETUP

µ

S

ppt

ppm

A

A

CAL

SETUP

µ

S

B

CAL

SETUP

C

C

B

Содержание 35631-00

Страница 1: ...me period please return freight prepaid and correction will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products w...

Страница 2: ...ady indicator auto endpoint function 28 P3 2 Selecting number of pH calibration points 29 P3 3 Selecting C or F 30 6 5 Program 4 0 Resetting to factory default settings pH 31 6 6 Program 5 0 Viewing p...

Страница 3: ...splay also shows error messages keypad functions and program functions mS ATC READY MEAS CAL HOLD ERR F SETUP ON OFF C S K pH mV pH ppt ppm 2 1 Primary display Secondary display 3 2 21 20 14 17 16 15...

Страница 4: ...ions In advanced set up mode Press CAL MEAS to return to main menu from sub menus Press CAL MEAS again to return to measurement mode from main menu ENTER Press to confirm values in Calibration mode an...

Страница 5: ...To connect the pH electrode 1 Slide the BNC connector of the probe over the BNC connector socket on the meter Make sure the slots of the connector are in line with the posts of the socket Rotate and...

Страница 6: ...eplaced on a point by point basis in pH and on a range by range basis in conductivity or TDS For example pH if you previously calibrated your meter at pH 4 01 7 00 and 10 01 and you recalibrate at pH...

Страница 7: ...each of the ranges you are measuring All new calibration data will over ride existing stored calibration data for each measuring range you calibrate If you are measuring in ranges near to or greater t...

Страница 8: ...and 1 If necessary press the MODE key to select TDS mode See figure 2 Rinse the probe with de ionized water or a rinse solution then rinse with a small amount of calibration standard 3 Dip the probe i...

Страница 9: ...nal reading To exit this program without confirming the temperature calibration value DO NOT press ENTER in step 7 Press CAL MEAS instead 5 Measurement To take readings 1 Rinse the pH electrode with d...

Страница 10: ...y to toggle between conductivity and pH readings Taking Conductivity or TDS Measurements ATC READY MEAS C S A A 5 2 B B Immerse probe beyond upper steel band Notes You can use the conductivity TDS man...

Страница 11: ...h time you turn the meter off 5 4 From measurement mode 1 Press the Mode key to select conductivity measurement mode 2 Press Setup key to enter Set Up mode 3 Press the and keys to scroll through subgr...

Страница 12: ...E C CAL F F The meter will now compensate readings for the manually set temperature Notes To exit this program without confirming the manual temperature compensation value DO NOT press ENTER in step...

Страница 13: ...onductivity TDS calibration data R2 R2 0 0 199 9 S 10 0 99 9 ppm P5 3 View conductivity TDS calibration data R3 R3 0 1999 S 100 999 ppm P5 4 View conductivity TDS calibration data R4 R4 0 00 19 99 mS...

Страница 14: ...electrode parameters for diagnostic purposes It lets you view P2 1 Electrode offset P2 2 Electrode slope SETUP pH A A SETUP mV pH B B 6 3 SETUP pH C C From measurement mode 1 Press the Mode key to se...

Страница 15: ...See figure 5 Press the and keys to select the configuration you require OFF switches the READY indicator off ON switches the READY indicator on ON and HOLD together switches the auto endpoint feature...

Страница 16: ...the ENTER key See figure 5 Press the and keys to toggle between NO and YES NO retains current settings YES resets to factory default settings 6 Press the ENTER key to confirm selection and to return...

Страница 17: ...to your calibration 6 7 mS SETUP S ppt ppm A A SETUP K B B From measurement mode 1 Press the Mode key to select conductivity or TDS measurement mode 2 Press the Set up key to enter Set Up mode See fig...

Страница 18: ...eter 7 1 See figure 5 Press the and keys to select the configuration you require ON switches the READY indicator on OFF switches the READY indicator off ON and HOLD together switches the auto endpoint...

Страница 19: ...solution See Appendix 2 on page 49 for information on how to calculate the TDS factor 7 Press the and keys to select your calculated TDS conversion factor 8 Press the ENTER key to confirm selection a...

Страница 20: ...e temperature coefficient is the amount of change in conductivity per degree of temperature it is expressed in percent per C or F Entering the exact temperature coefficient of your solution lets you a...

Страница 21: ...ode into the heated solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposits prepare a 1 pepsin solution in 0 1 M of HCl Set the electrode in the soluti...

Страница 22: ...nstead of performing this reactivation procedure 1 Dip or stir the pH electrode in alcohol for 5 minutes 2 Leave the electrode in tap water for 15 minutes 3 Dip and stir the electrode in concentrate a...

Страница 23: ...al LCD Operating temperature 0 to 50 C 32 to 122 F Power four 1 5 V AAA batteries included Battery life 200 hours continuous use Probe connectors pH mV BNC connector Conductivity TDS Temperature notch...

Страница 24: ...00654 00 pH 4 01 calibration buffer 1 pint WD 00654 04 pH 7 01 calibration buffer 1 pint WD 00654 08 pH 10 01 calibration buffer 1 pint WD 00654 12 pH 12 45 calibration buffer 1 pint Model No Solution...

Страница 25: ...in the above formula The result should be the TDS value 12 Appendix 1 Conductivity to TDS Conversion Factors 1 Factor the conductivity to ppm TDS conversion factor Multiply conductivity by this factor...

Страница 26: ...the meter readings 15 Appendix 4 Meter factory default settings Resetting the meter to factory default settings clears all calibration data and most other setup functions you might have changed The fo...

Отзывы: