P7.2 Selecting °C or °F
This meter lets you select between °C and °F units for temperature readings.
From measurement mode
1.
Press the Mode key
to select
conductivity or TDS measurement mode.
2.
Press Setup key
to enter Set Up mode.
3.
Press the
▲
and
▼
keys
to scroll through
subgroups until you view parameter P7.0.
See figure
4.
Press the ENTER key
four times to select
parameter 7.2.
See figure
5.
Press the
▲
and
▼
keys
to toggle between
°C and °F.
6.
Press the ENTER key
to confirm selection
and to proceed to step 4 of P7.3. Press the
CAL/MEAS key twice to return to meas-
urement mode.
mS
SETUP
µ
S
ppt
ppm
C
C
D
Notes
You can also switch between °C and °F in Program P3.3 (available from pH mode).
If you switch between °C and °F in conductivity mode, the meter will also switch in
pH mode.
C
SETUP
F
SETUP
D
From measurement mode
1.
Press the Mode key
to select
conductivity or TDS measurement mode.
2.
Press Setup key
to enter Set Up mode.
3.
Press the
▲
and
▼
keys
to scroll
through subgroups until you view
parameter P7.0.
See figure
4.
Press the ENTER key
to select
parameter 7.1.
See figure
5.
Press the
▲
and
▼
keys
to select the
configuration you require.
•ON switches the READY indicator on.
•OFF switches the READY indicator off.
•ON and HOLD together switches the
auto endpoint feature on.
6.
Press the ENTER key
to confirm selection
and to proceed to step 4 of P.7.2. Press the
CAL/MEAS key twice to return to meas-
urement mode.
P7.0: Conductivity/TDS measurement configuration
P7.1: READY indicator and auto endpoint function
This program lets you select:
• “READY indicator on” to indicate when the reading is stable.
• “READY indicator off” for faster meter response.
• Auto endpoint function on. Select auto endpoint on to “hold” the reading when
it is stable for more than 5 seconds. The display automatically freezes, and the
HOLD indicator appears on the left side of the display. Press the HOLD key to
release the display and access other functions.
mS
SETUP
µ
S
ppt
ppm
READY
SETUP
ON
READY
SETUP
OFF
READY
HOLD
SETUP
ON
6.8
A
A
B
B
Notes
Meter default is set for Ready indicator on,
and auto endpoint function off.
You can also change the Ready indicator
and auto endpoint function in Program P3.1
(available from pH mode). Any changes
you make to the Ready indicator/ auto end-
point function in conductivity mode will
also change in pH mode.