OM0232
|
Rev 7 July/2017
Page 21 of 56
NuAire, Inc.
|
2100 Fernbrook Lane
|
Plymouth, MN 55447
|
U.S.A
|
ph: 763.553.1270
|
fx: 763.553.0459
|
tf: 800.328.3352
|
www.nuaire.com
6.2.7 Interrupt centrifugation
Visual
Instruction
Control
During a centrifugation, press the Start/Stop
button.
The rotor will start slowing down until
complete stop
Open the lid.
6.2.8 Open after an electric power failure / Emergency Stop
This operation must be performed by a person trained in
centrifugation-related risks.
In case of power failure, the lid cannot be opened by the engine
unlocking control.
You must wait 15 minutes for the rotor to come to a complete stop
before opening manually.
Turn off the power by pressing the switch to the “0”
position.
Punch the safety label on the front panel, using the
wrench provided. Insert it fully into the cavity.
Turn it clockwise to release the lid.
Remove the wrench.
Unload samples.
Following a manual release, the wrench must be removed from its housing to prevent accidents related to the
rotation of the wrench during the activation of the locking mechanism. In addition, the security indicator label must
be replaced during a safety check carried out by an authorized technician.
CAUTION
!
2
3
1