6.1.4 Using the interface
6.1.5 Opening the lid
On the first start, the lid may not open right away due to a long storage in a cold environment.
In this case, warm-up the jacks by repeated manual shifting of the lid downwards.
To open the lid, press the button
.
6.1.6 Rotor loading and balancing
Position the 4 numbered buckets in their corresponding positions
on the rotor.
Each bucket should pivot freely in its slot.
Place the inserts corresponding to the tubes to be centrifuged in
the four buckets
The buckets must be symmetrically distributed.
For better balance, each numbered bucket must be placed in the
corresponding position marked on the rotor.
Each position on the rotor must be equipped with its bucket.
Do not use a rotor without its complete set of four buckets.
Maximum loads according to the type of rotor are specified in the
annex.
The centrifuge tolerates an imbalance of 15grams.
If the imbalance is higher than the tolerated threshold, an
imbalance detection system stops the centrifuge automatically.
The following message appears: "Imbalance" and the rotor stops in
free wheel mode without braking system.
Excessive imbalance is likely to damage the turning components
and/or the centrifuge.
Loading the buckets
Allowed:
Forbidden :
Loading the rotor
Allowed:
Symmetrical loading
Forbidden:
Incomplete loading
Asymmetrical loading
1.
Touch screen for direct access to various functions.
2.
Single button for easy operation.
This button allows the validation, the start of a cycle, but also
the interruption of a cycle, and the output of a menu.
The functions accessible through this single button are
activated as required by the microcontroller.
2
1