OM0301
August 20
Page 58 of 105
Step 3: Place a velometer in the cabinet workzone on the horizontal plane 4 inches (102mm) above the
bottom edge of the window. Spot check several points on the recommended downflow velocity
Step 4: Press + or – button of supply duty cycle to adjust blower speed.
The objective of this spot check is to obtain the desired downflow average velocity as close as
possible to the stated goal of 60 LFPM (.30 m/s).
DON'T SPEND MORE THAN 5 MINUTES SPOT CHECKING.