OM0301
August 20
Page 43 of 105
When installing new filters,
USE ONLY NUAIRE SPECIFIED FILTERS FOR REPLACEMENT.
7.3 HEPA Filter/Motor Replacement
The HEPA Filters under normal usage and barring an accident (a puncture), do not need replacement until the exhaust
volume cannot be maintained or the access inflow velocity cannot be maintained at 100 LFPM (min.) .51 m/s). This may
permit the average downflow velocity to be as low as 55 LFPM (.28 m/s) as long as no point falls below 20 percent of the
average downflow velocity.
The HEPA Filters should not be replaced until the entire cabinet has been decontaminated or known to be biologically
"clean". Constant pressure spring-type clamps are used to hold the exhaust filter tightly in place to counteract seal
relaxation, while the supply filter employs Nuaire's HEPEX pressure plenum.
USE ONLY REPLACEMENT FILTERS OF THE
SAME RATED FLOW AND SIZE AS ORIGINALLY INSTALLED, TO INSURE PROPER AIRFLOW BALANCE CAN BE ACHIEVED.
It is not always necessary to replace both the supply and exhaust filters at the same time. In fact, it is highly likely that the
exhaust filter will need replacement far more often than the supply filter, due to (1) the larger volume of air passing
through it, (2) it’s much smaller size, and (3) the capability of the exhaust system.
7.3.1 Supply Filter Replacement
(see Drawing BCD-19032)
Disconnect electrical power from the cabinet before attempting any maintenance action.
Step 1: Remove screws at each upper side of the control center and allow the control center to rotate down,
resting on the safety straps. Second, remove the front decorative panel, which is held into position by (3)
screws on the top edge.
Step 2: Place sliding window into lowest position and remove front filter panel, which is held into position by
Phillips pan head screws. Once the screws are removed, remove the panel.
NOTE: Screws are used in lieu of acorn nuts, and lock washers.
The screws have O-rings and should be replaced if damaged or badly deformed.
Step 3: Remove filter access panel, which is held into position by screws. Remove screws and the panel.
The interior of the cabinet is now fully exposed for replacement of the filter.
Description:
Supply HEPA Filter
Exhaust HEPA Filter (UL 586)
Efficiency:
99.99% @ 0.3 Micron
99.99% @ 0.3 Micron
Airflow Rating:
100 fpm @ .55
±
.05" w.g. per sq. ft.
250 fpm @ .45
±
.05" w.g. per sq. ft.
Frame Type:
Metal
Metal
NU-565-400
Nuaire Part Number: A-980979-01
A-980845-03
Filter Size:
21" (533mm) x 44" (1118mm) x 3" (76mm) 20" (508mm) x 22” (559mm) x 11 1/2" (292mm)
Filter Manufacturer: Camfil Farr/FG/AAF
Camfil Farr/FG/AAF
NU-565-600
Nuaire Part Number: A-980979-03
A-980845-04
Filter Size:
21" (533mm) x 68" (1727mm) x 3" (76mm) 20" (508mm) x 36" (914mm) x 11 1/2" (292mm)
Filter Manufacturer: Camfil Farr/FG/AAF
Camfil Farr/FG/AAF
Содержание LabGard ES NU-565-400
Страница 7: ...OM0301 August 20 Page 7 of 105...
Страница 9: ...OM0301 August 20 Page 9 of 105...
Страница 10: ...OM0301 August 20 Page 10 of 105...
Страница 11: ...OM0301 August 20 Page 11 of 105...
Страница 17: ...OM0301 August 20 Page 17 of 105...
Страница 18: ...OM0301 August 20 Page 18 of 105...
Страница 45: ...OM0301 August 20 Page 45 of 105...
Страница 50: ...OM0301 August 20 Page 50 of 105...
Страница 82: ...OM0301 August 20 Page 82 of 105...
Страница 98: ...OM0301 August 20 Page 98 of 105...
Страница 99: ...OM0301 August 20 Page 99 of 105...
Страница 100: ...OM0301 August 20 Page 100 of 105...
Страница 101: ...OM0301 August 20 Page 101 of 105...
Страница 102: ...OM0301 August 20 Page 102 of 105...
Страница 103: ...OM0301 August 20 Page 103 of 105...
Страница 104: ...OM0301 August 20 Page 104 of 105...
Страница 105: ...OM0301 August 20 Page 105 of 105...