OM0301
August 20
Page 49 of 105
Press SET MOTOR TYPE to verify correct setting.
Upon a MASTER RESET, the motor type is defaulted to DC ECM Single.
The NU-565 Series 1 (115Vac) requires the motor type selected to be DC ECM Single.
Always verify motor type when verifying cabinet type.
7.5.2.2 Setpoints/Limits
The airflow setpoints and alarm limits may also be verified or altered. Typically, these default values are
factory set based on the cabinet type, model and size as previously discussed. However, they may be altered
in special cases for modified cabinets. The setpoint establishes the airflow values that are to be maintained
The high low limits establish the alarm boundaries from the nominal setpoint. The default values have been
established based upon the performance specifications and cabinet component tolerances.
Entry into the setpoints/limits functions requires a service password for entry. The default password is “9876”
Содержание LabGard ES NU-565-400
Страница 7: ...OM0301 August 20 Page 7 of 105...
Страница 9: ...OM0301 August 20 Page 9 of 105...
Страница 10: ...OM0301 August 20 Page 10 of 105...
Страница 11: ...OM0301 August 20 Page 11 of 105...
Страница 17: ...OM0301 August 20 Page 17 of 105...
Страница 18: ...OM0301 August 20 Page 18 of 105...
Страница 45: ...OM0301 August 20 Page 45 of 105...
Страница 50: ...OM0301 August 20 Page 50 of 105...
Страница 82: ...OM0301 August 20 Page 82 of 105...
Страница 98: ...OM0301 August 20 Page 98 of 105...
Страница 99: ...OM0301 August 20 Page 99 of 105...
Страница 100: ...OM0301 August 20 Page 100 of 105...
Страница 101: ...OM0301 August 20 Page 101 of 105...
Страница 102: ...OM0301 August 20 Page 102 of 105...
Страница 103: ...OM0301 August 20 Page 103 of 105...
Страница 104: ...OM0301 August 20 Page 104 of 105...
Страница 105: ...OM0301 August 20 Page 105 of 105...