Recommended separation distances between portable and mobile RF communications equipment and the
device
Rated maximum output
power of transmitter W
0.01
0.1
1
10
100
0.12
0.37
1.17
3.69
11.67
0.12
0.37
1.17
3.69
11.67
0.23
0.74
2.33
7.38
23.33
Separation distance according to frequency of
transmitter m
150 kHz to
80 MHz
d=1.2√P
80 MHz to
800 MHz
d=1.2√P
800 MHz to
2.7 GHz
d=2.3√P
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE I At 80 MHz and 800 MHz. the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
people.
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and as
recommended below, according to the maximum output power of the communications equipment.
TABLE 4