19
Conducted RF
lEC 61000-4-6
Radiated RF
lEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
3 Vrms
150kHz to
80 MHz
10 V/m
80 MHz to
2.7 GHz
3 Vrms
150kHz to
80 MHz
Portable and mobile RF communications
equipment should be used no closer to any part
of the UT1032 device, including cables, than the
recommended seperation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended seperation distance:
d=1.2√P
d=1.2√P , 80MHz to 800MHz
d=2.3√P , 800MHz to 2.7GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation
distance in meters (m).
determined by an electromagnetic site survey
a
,
should be less than the compliance level in each
frequency range
b
.
Interference may occur in
the vicinity of equipment marked
with the following symbol:
NOTE I: At 80 MHz ends 800 MHz the higher frequency range applies.
a
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
the location in which the device is used exceeds the applicable RF compliance level above, should be
observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the device.
b
i
]V/m.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Guidance and manufacturer’s declaration - electromagnetic emissions
the user of the UT1032 should assure that it is used in such an environment.
TABLE 3
UT1032
19