14
UNIT SPECIFICATIONS
MODULATION
DUTY FACTOR
5%
50%
100%
WAVE CHARACTER
Low
Medium
High
OUTPUT POWER
0.32W±20%
3.20W±20%
6.40W±20%
PROGRAM
L
M
H
Ultrasound
Probe
Ultrasound Modulation Frequency:
Max. Output Power:
Output Power:
Pulse Repetition Rate:
Modulation Duty Cycle:
Waveform:
R
BN
(Max):
Working Time:
Preheat Temperature:
Preheat Time:
Dimension:
Weight:
Material of Applicator:
Beam Type:
Degree of Protection against Water:
Input:
Output:
Time:
Mode:
Time Indication Lights:
Duty Cycle Indication Lights:
Item Description
1.0MHz±10%
6.4W±20% (Modulation duty cycle at100%)
L: 0.32W±20% M: 3.20W±20% H:
6.40W±20%
100Hz±10%
L (5%), M (50%), H (100%)
4.0cm²± 20%
Pulsed, Continuous
5.0
1.6W/cm
2
±20% (Modulation duty scycle at 100%)
L: 0.08W/cm
2
±20% M: 0.80W/cm
2
±
H:1.60W/cm
2
±20%
Adjustable at 5 minutes, 10 minutes, 15 minutes
Max. 35±5 degree centigrade
(NOTE: Actual preheat temperature
will be influenced by the environmental temperature and preheat time.)
Max. 3 minutes
202 mm (L) x 49 mm (W) x 70 mm (H)
193g (without adapter)
Aluminum Alloy
Collimated
IPX7 (Only for Treatment Head)
Voltage: AC 100-240V, 50/60Hz, 0.5A
Output voltage: DC 15V, Max. Currency: 1.2A
Choose working time: 5min — 10min —15min
—0min (stop)
Choose modulation duty cycle: 5% —50% — 100%
5, 10, 15 minutes
Low (L), Medium (M), High (H)
Power
Adapter
Indication
Lights
Buttons
Program
Lists
Operating Conditions: 5°C ~ 40°C; 30% ~ 75%RH; 800~1060hPa
Service life: 2 years; Safety class: Class II,type BF
Beam type:
Pulse duration:
collimated
0.5ms,5ms,10ms
Содержание UT1032
Страница 1: ...INSTRUCTION MANUAL UT1032 Portable Ultrasound Device www nutekmedical com ...
Страница 22: ......