12
1. The device has a
load detection system
for safety. When the treatment
head does not have good contact with the skin, the device will stop
treatment automatically, and the time indicator light will flash one time. The
device will not continue the treatment program until good contact is made.
2. The device has a temperature protection function. When the temperature
of the treating head exceeds 107°F (42°C), the treatment will automatically
stop and the time indicator light will flash two times. The device will not
continue the treatment program until the temperature is below 104°F (40°C).
LOAD DETECTION SYSTEM CAUTION:
MAINTENANCE
be cleaned with a damp cloth. Use lukewarm water and a non-abrasive liquid
household cleaner (no abrasive, no alcohol content solution). If a more sterile
cleaning is needed, use a cloth moistened with an antimicrobial cleaner.
CAUTION:
Do not submerse the device in liquids. Should the unit accidentally
become submersed, contact the dealer or Authorized Service center
immediately. Do not attempt to use the device that has been submersed in
any liquid substrate until inspected and tested by a Service Technician
Cleaning of the applicator
The applicator should be regularly inspected for damage, e.g. hairline cracks,
which could allow the penetration of liquids. Clean the contact surface
immediately after each treatment. Make sure that no ultrasound gel remains
on the applicator. We further recommend cleaning the head, cable and
adapter daily, using a soft cloth damped with lukewarm water. The applicator
can be disinfected using a cloth moistened with an antimicrobial cleaner.