For transmitters rated at a maximum output which is not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable to
the respective column, whereby P is the maximum rated output of the transmitter in Watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz the higher frequency range is applicable.
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
G ¥3
80 MHz to 800 MHz
G ¥3
800 MHz to 2.5 GHz
G ¥3
Rated output of transmitter
W
0.01
0.035
0.12
0.07
0.22
0.38
1.2
0.35
0.70
0.11
1.1
2.2
12
3.5
3.8
7
0.1
1
10
100
86
Rehab-Series