Cycle Time........................................continues,10/10,10/20,10/30,10/50,custom:0-60
Ramp.........................................................................0/0,1/1, 2/2,5/5,8/8,custom:0-9
10.2.8 High Voltage
The High Voltage Pulsed Current (HVPC) has a very brief pulse duration characterized by two
distinct peaks delivered at high voltage. The waveform is monophasic (current flows in one
direction only). The high voltage causes a decreased skin resistance making the current
comfortable and easy to tolerate.
Frequency...................................................................................1-120Hz,stepping 1Hz
Polarity................................................................................................Positive,negative
P.Dur.................................................................................................................100us
Intensity.....................................................................................CV:0-500V,stepping 5V
Channel mode....................Single channel Mode,Reciprocal Mode,Co-Contraction Mode
Treatment Time.........................................................................................1-60 minutes
Cycle Time........................................continues,10/10,10/20,10/30,10/50,custom:0-60
Ramp.........................................................................0/0,1/1, 2/2,5/5,8/8,custom:0-9
10.2.9 NMS
10.2.9.1NMS
Mode...................................................................................................NMS,NMS Burst
Frequency...................................................................................1-250Hz,stepping 1Hz
P.Dur....................................................................................20us-1000us,stepping 5us
Intensity.....................................CC:0-200mA,stepping 0.5mA;CV:0-200V,stepping 0.5V
Channel mode....................Single channel Mode,Reciprocal Mode,Co-Contraction Mode
Treatment Time.........................................................................................1-60 minutes
Cycle Time........................................continues,10/10,10/20,10/30,10/50,custom:0-60
Ramp.........................................................................0/0,1/1, 2/2,5/5,8/8,custom:0-9
10.2.10 Galvanic
10.2.10.1 Continuous
Polarity................................................................................................Positive,negative
Intensity................................................CC:0-80mA,stepping 5uA;CV:0-80V,stepping 5V
79