The lifetime of the instrument is determined by the number of charging cycles. The internal batteries can be fully charged more than 500 times before they need to
be replaced. The instrument should not be left uncharged for more than 1 year.
The MulTipeg Driver is guaranteed for at least 100 autoclave cycles, and a MulTipeg is guaranteed for at least 20 autoclave cycles, before they are degraded in any
way.
17. Troubleshooting
17.1 Possible errors
Difficult to achieve a measurement:
In some cases it is more difficult for the instrument to bring the MulTipeg into vibration. If so, try to hold the instrument tip closer to the tip of the peg. Check
also that no soft-tissue is touching the peg which might stop its vibration.
Noise warning (audible and visible on the display):
An electric device close to the instrument is causing the warning. Try to remove the source.
The instrument suddenly turns off:
The instrument turns off automatically after 1 minute of inactivity. It also turns off if the battery level is too low, and due to any of the error codes described
below.
Not all segments are lit up when instrument is started:
The instrument is damaged and has to be sent for repair.
17.2 Error codes
If malfunctioning, these error codes are shown on the display before it turns off:
E1: Hardware error. Malfunctioning electronics
E2: Noise error. Shown if constant electromagnetic noise is present
E3: Pulse power error. Malfunctioning magnetic pulse generation
18. Accessories & Spare Parts
Model
MulTipeg Driver
Sterile Cover
Mains adapter
EU plug
UK plug
AU plug
US plug
REF
55003
55105
55093
incl 1.5 m cable
55094
55095
55096
55097
MulTipeg: Please refer to the updated list from the supplier.
19. Service
In case of a malfunctioning instrument, the device should be sent to the manufacturer or distributor for repair or exchange.
Osseo100 is covered by a two-year warranty.
Use of accessories other than those specified or provided by the manufacturer of this equipment could result
in increased emissions or decreased electromagnetic immunity of this equipment and result in improper
operation.