F type Spray Nozzle
Spray Lubricant (Option)
Contra Angle Head
Connector
Connector
14
15
ENDO-MATE DT uses rechargeable batteries. They can be recharged 300-500 times, depending on the use conditions of
the Control Unit. If the operating time or recharging time becomes shorter or the rotation power becomes weaker, although
the “memory effect” described in “ NOTICE” is not applicable, the life of the batteries is suspected. In such a case, ask
your dealer to replace the batteries or replace them with new ones by yourself. (Refer to the “ Changing Batteries.”)
When replacing them by yourself, be sure to observe the following “ CAUTIONS ON CHANGING BATTERIES.” Note that
NSK shall not be held liable for any malfunction or failure resulting from your not following the “ CAUTIONS ON
CHANGING BATTERIES.”
* we recommend our geunine rechargeable Batterie.
Order Code : U421070
Changing Batteries
1) Turn off the power. Remove the AC Adaptor.
2) Remove the batteries after sliding the battery cover on the
back of the control unit in the direction of the arrow. (Fig. 13)
3) Take out the old batteries, and pull out the cord, holding it
at the connector. (Fig. 14)
4) Insert the connector of the cord of new batteries into the
connector in accordance with the polarity indication label
inside the battery chamber, and place the batteries into
the chamber with care so as not to catch the cord.
9. CHANGING BATTERIES
In case of difficult to insert the connector onto the battery, the polarity may be incorrect. Do not insert it by force.
CAUTION
Make sure to remove the AC Adaptor before
changing batteries. Otherwise, it may cause a
malfunction.
CAUTION
Make sure to pull out the cord by holding the
connector. Otherwise, it may cause a break in the cord.
CAUTION
• Do not open any part other than the battery cover.
• Be sure to purchase and use only the recommended batteries. Otherwise, batteries may cause damage, fluid
leakage or explode.
• Do not work with wet hands. This may cause trouble due to short-circuiting of batteries and moisture infiltrating
this product.
CAUTIONS ON CHANGING BATTERIES
Autoclaving
1) Brush the dirt off the surface of the contra angle head, and wipe it off with a cotton cloth moistened with rubbing
alcohol. Do not use metal brush.
2) Lubricate the head with the spray lubricant (option).
(Refer to the “P.14 7. (1) Lubricating Contra Angle Head.”)
3) Insert the head into an autoclave pouch and seal it.
4) Autoclavable up to max. 135°C.
ex.) Autoclave for 20 min. at 121°C, or 15 min. at 132°C.
5) Keep the contra angle head in the autoclave pouch to keep it clean until you use it.
* Sterilization at 121°C for more than 15 minutes is recommended by EN13060 or EN ISO17665-1.
• Sterilize the contra angle head only.
• For the sterilization method, we recommend the autoclave sterilization method.
• Sterilization is required first time you use and after each patient as noted below.
8. STERILIZATION
Do not apply high-temperature sterilization methods such as autoclave sterilization to any part other than the
contra angle head.
CAUTION
(1) Lubricating Contra Angle Head
• Lubricate the contra angle head only.
• Supply PANA SPRAY/PANA SPRAY Plus (option) after each
use and/or before each calibration and autoclaving.
1) Screw the F type spray nozzle onto the spray nozzle by
approx. 10 turns.
2) Insert the F type spray nozzle into the rear part of the
contra angle head and lubricate the head for 2-3
seconds. (Fig. 12)
(2) Cleaning Motor Handpiece
When the motor handpiece becomes dirty, wipe it off with a cotton cloth moistened with rubbing alcohol.
7. CLEANING
• Do not lubricate the motor section of the motor handpiece.
• To clean the motor handpiece, do not use any solvent such as benzine and, thinner.
• Before mounting the lubricated contra angle head to the motor handpiece, wipe off extra oil. Stand on its end or
lean it in the proper position for gravity draining. Mount it after excess oil has been drained.
• Hold the contra angle head securely to prevent it from flying off by the pressure of the spray.
• Supply lubricant until it comes out of the handpiece head (for approx. 2 seconds).
• Do not use a spray can upside down. In such a case, only spray gas comes out and oil is not sprayed.
CAUTION
(5) Alarm Sound Volume Control
The alarm sound volume can be controlled at the three steps of “BIG VOLUME, SMALL VOLUME and LIMITED OFF. ”
1) Keep pressing both buttons of the PROGRAM key simultaneously.
2) The alarm mark on the LCD panel changes, and the sound volume changes.
3) Release the PROGRAM key when the sound volume reaches your required level and the required sound level is set up.
• The sound level is retained, even if the power is turned off.
• If P.12 6. (2) “Initialization of Program (Factory-set configuration)” is executed, the sound is set to “BIG VOLUME.”
NOTICE
Fig. 12
* The contra angle head in the figure is MP-F20R.
Fig. 13
Battery Cover
Battery Cover
Fig. 14