(4) Calibration
This function is to decrease fluctuation in the rotation speed of the motor handpiece and the difference in torque by the
contra angle head.
1) Lubricate the contra angle head with the PANA SPRAY/PANA SPRAY Plus (option). (Refer to the “P14 7. (1) Lubricating
Contra Angle Head.”)
2) Turn off the power if the power is on.
3) Connect the AC Adaptor and confirm that the AC IN lamp lights.
4) Turn on the power.
5) Hold down both buttons of the SPEED key simultaneously for more than one second.
6) The LCD panel displays “ ” with the alarm sound.
7) Attach the contra angle head to the motor handpiece and press the MEMORIZE key.
8) If the motor handpiece begins to rotate, leave it as it is until it stops.
9) This process ends, if the rotation stops and the LCD panel display returns to its original state.
10) If you want to stop this process, turn off the power.
(3) Battery Refresh
Nickel-metal-hydrate batteries may result in a decrease in charging capacity, if additional charging (repeating a cycle of
short-time use and recharging) is made. (Generally, this phenomenon is called a “memory effect.”) This Battery Refresh
function is to resolve this phenomenon.
1) Turn off the power if the power is on.
2) Connect the AC Adaptor and confirm that the AC IN lamp lights.
3) Hold down the POWER key for more than one second, while pressing the REVERSE key.
4) The alarm sounds for a length of time, and the refresh mode is activated. At this time, the mark of the remaining amount
of the batteries on the LCD panel is animated in the direction opposite to display at the time of charging. At the same
time, the CHARGE lamp flashes slowly.
5) Leave it as it is. The batteries are discharged and recharged automatically.
6) Hold down the POWER key for a while, when you want to stop this process. (At this time, the power does not turn on.)
• Lubricate the contra angle head before Calibration.
• This function does not operate if AC Adaptor is not used.
• Do not touch or apply a load to the rotating shaft of the contra angle head. It obstructs accurate measurement.
• This function cannot completely absorb the individual difference of the actual motor handpiece and contra angle head.
NOTICE
• This function is not activated unless this product is powered by the AC Adaptor.
• If this function is used, all programs will disappear and return to the originally set values. Record the present
program details according as needed.
NOTICE
• This function is not activated, if not driven by the AC Adaptor.
• It takes about maximum of 5 hours to discharge batteries.
• It takes 5 hours to discharge and 5 hours to recharge. Therefore, total of about 10 hours are required. For
example, if dental treatment is finished at 7 p.m. and the refresh mode is set, the batteries will be fully charged
by 5 a.m. the next morning at the longest.
• This function is not required for each charging. Try this function, if the operating time is shortened even if the
batteries are relatively new.
• Do not repeat activation and deactivation of the refresh mode in a short time. It may result in enhancement of
the memory effect.
• This function is fully effective for solution of a “memory effect.” However, it cannot be solved completely at one
time due to characteristics of batteries. We recommend you repeat this process a few times according to need.
NOTICE
(1) Program
You can change any preset values and have them memorized to your desired settings (rotation speed, torque limit value,
gear ratio and auto reverse mode).
1) Press the PROGRAM key until it turns to the program number which you want to have memorized.
2) Adjust the rotation speed, torque limit value, gear ratio and auto reverse mode by each key according to your needs.
3) Hold down the MEMORIZE key for more than one second. When the alarm sounds, memorization is completed.
(2) Initialization of Program (Factory-set configuration)
The program can be returned to the original state configured at the time of factory shipment, if setup becomes confusing.
1) Turn off the power, when the power is on.
2) Connect the AC Adaptor and confirm that the AC IN lamp lights up.
3) Hold down the POWER key for more than one second, while pressing the AUTO REVERSE key and MEMORIZE key at the
same time.
4) Release the keys when the LCD panel indicates “ ”, and re-press the MEMORIZE key while this is displayed.
5) Initialization is completed, when “ ” is displayed.
6. CONVENIENT FUNCTIONS
• The program cannot be memorized while the motor handpiece is in motion.
• The rotational direction is not memorized in the program. The program cannot be memorized if the rotational
direction is set to the reverse rotation (R).
• If the PROGRAM key is kept pressed, it is fast-forwarded.
• The program is not memorized unless the MEMORIZE key is held down for more than one second. If the
program number is changed by the PROGRAM key without pressing the MEMORIZE key, the initially memorized
preset values remain. (Cancel function)
NOTICE
(6) Completion of Medical Treatment
When the treatment is completed, return the motor handpiece to the handpiece stand and hold down the POWER key for
more than one second to turn off the power.
If 10 minutes pass without operation of the ON/OFF button and Foot Control (Option), the power turns off
automatically. (Auto power off function) However, the power does not turn off automatically if the motor handpiece
is in motion, even if it is not used.
NOTICE
AUTO REVERSE OFF
The motor handpiece stops without reverse rotation. In this instance, the LCD panel shows “ ” and the rotation
speed alternately.
• When the reverse rotation (R) is set, this Auto Reverse Function is not activated.
• When the batteries go down (when the mark of the remaining amount of the batteries indicates “ , ”),
the actual load may not reach the preset torque limit value. In this case (in the case of operation by batteries),
this auto reverse function will not be activated. When high torque is required, use the AC Adaptor or use this
product in a condition where the batteries are not significantly consumed (when the mark of the remaining
amount of the batteries indicates “ , ”).
• If a load is continuously applied to the motor handpiece, it may automatically stop to prevent overheating. (The
torque limit value indicates “ ” on the LCD panel). In this case, leave the motor handpiece for a while until
it cools down.
When the display returns to normal, the motor handpiece can be used.