Classifications of equipment
• Type of protection against electric shock :
– Class II equipment :
• Degree of protection against electric shock :
– Type B applied part :
• Method of sterilization or disinfection recommended by the manufacture :
– See 8. Sterilization
• Degree of protection against ingress of water as detailed in the current edition of IEC 60529 :
– Foot Control (Option) : IPX1 (Protected against vertically falling water drops)
• Degree of safety of application in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous
oxide :
– EQUIPMENT not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous
oxide.
• Mode of operation :
– Continuous operation
00
1
Thank you for purchasing the ENDO-MATE DT.
This product is used in dental office only.
This product is used primarily for root canal enlargement. Read this Operation Manual carefully before use for operation
instructions, care and maintenance. Keep this Operation Manual for future reference.
Read these safety cautions thoroughly before use and operate the product properly.
These indicators are to allow you to use the product safely and prevent danger and harm to you and others. These are
classified by degree of danger, damage and seriousness. All indicators concern safety, and always follow them.
Cautions for handling and operation
CONTENTS
Cautions for handling and operation ················1
1. FEATURES ·······················································3
2. SPECIFICATIONS ··············································4
3. NAME AND FUNCTION OF EACH PART ··············4
4. HOW TO CONNECT EACH PART ························8
5. HOW TO USE ···················································9
6. CONVENIENT FUNCTIONS ······························ 12
7. CLEANING ····················································· 14
8. STERILIZATION ·········································· 14
9. CHANGING BATTERIES ······························ 15
10. ERROR CODE ··········································· 16
11. TROUBLESHOOTING ································· 17
12. CONTRA ANGLE HEAD (OPTION) ················ 18
13. WARRANTY ·············································· 19
14. DISPOSING PRODUCT ······························· 19
D A N G E R
WA R N I N G
CAUTION
N O T I C E
Explains an instruction where death or serious injury may occur.
Explains an instruction where bodily injury or damage to device may occur.
Explains an instruction where possibility for minor to medium bodily injury or damage to device
may exist.
Explains an instruction that should be observed for safety reasons.
Classification
Degree of danger or damage and seriousness