8.4
Transfer characteristics
The characteristic curves shown in the measurement report are created under the stated
environmental conditions and with the stated target material. When you order a complete
displacement sensor system, you will receive an individual measurement report which may deviate
from the illustrated typical transfer function.
8.4.1
Characteristic curve for distance measurement [GAP / V]
The transfer characteristic describes the interdependence of signal voltage and the distance
between sensor tip and the object to be measured.
•
Y-axis: ...................... GAP [V] ..... .................... X-axis : ...... Displacement [mm]
8.4.2
Characteristic curve for vibration measurement [ISF / mV/
μm]
The displacement sensor system has a sensitivity of -4 mV/
μm (-101,5 mV/mil) with reference to
B&K Vibro reference material 42CrMo4 (Material-No. 1.7225) according to DIN 17 200, corresponding
to AISI/SAE 4140.
The ISF (Incremental Scale Factor) describes the accuracy of the
sensitivity (ISF error / %) in dependence on the separation.
•
Y-axis: ...................... ISF [mV /
μm] ................. X-axis : ...... Displacement [mm]
8.4.3
Deviation from best fit straight line [DSL /
μm]
DSL (DSL/
μm = Deviation from best fit Straight Line) describes the deviation from an ideal
characteristic curve of the distance measurement.
•
Y-axis: ..................... DSL-error [
μm] ............... Y-axis: : Displacement [mm]