4.5
Electrical connection
Connect the driver with an electronic monitoring system (e.g. VC-6000) according to the drawing
below.When setting up your measuring system you should observe the following behavior of the
driver.
The heating effect of the driver electronics
During the warm-up phase the heating effect causes a drift in the static distance measurement.
Displacement sensor
system
Operating time
Drift
5 m
≥1 h
≤6 μm
10 m
≥1 h
≤12 μm
25 m
≥1 h
≤24 μm
Four wire connection
Figure 4-1)
Connection of the driver to a monitoring system (VC-6000).