background image

 

LogBox-DA Datalogger 

NOVUS AUTOMATION 

5/7 

INPUT CONNECTIONS – CHANNEL 1  

The  following  figures  show  connections  for  the  various  input  types 
accepted in channel 1. No jumper is required to adjust the channel 1 
input  type  (the  circuit  is  reconfigured  automatically  according  to  the 
input type selected). 

INPUT CONNECTIONS – CHANNEL 2 

For  adequate  use  of  channel  2,  set  internal  J1  and  J2  positioning 
according  to  the  input  signal  chosen,  as  shown  in  the  table  below 
(factory default is 4-20mA / 0-20mA signals). 

Input signal

 

J1 Position

 

J2 Position 

4-20 mA / 0-20 mA 

A

 

0-10V

 

B

 

0-50 mV

 

C

 

Table 1 – J1 and J2 positioning 

 

The following figures show connections for different types of signals 
applied to channel 2.   

 

EXTERNAL BATTERY SWITCH 

The  example  below  shows  the  usage  of  the  external  battery  switch 
for commanding the power supply of external devices. 
- Channel 2 is configured to 4-20mA input signal. A battery is used to 
provide power to the 4-20 mA loop. The battery switch “turns on” the 
power to the loop a moment (defined in the configuration) before the 
measurement  is  taken,  enabling  the  transmitter  (pressure, 
temperature, etc) to start up and stabilize the output. This approach 
saves  the  battery  energy,  as  the  power  is  turned  on  only  when 
required (before each measurement). 

Fig. 13

 – Example of the battery switch powering a transmitter. 

 

INSTALLATION RECOMMENDATIONS 

  Signal  wires  should  be  installed  in  grounded  conduits  and  away 

from power or contactor wires.  

  Installing  RC  filters  (47

  and  100nF,  serial)  is  strongly 

recommended at contactor coils or any other inductors. 

  System  failure  should  always  be  taken  into  account  when 

designing  a  control  panel  to  avoid  irreversible  damage  to 
equipment or people. 

OFFLOADING AND DATA VISUALIZATION

 

TRANSFERING RECORDS 

Transferring  the  data  to  a  PC  is  accomplished  by  clicking  on  the 

button 

 in the LogChart II software. The IR wand must be pointed 

to the logger during the process. Data can be collected anytime and 
saved in files for future analysis (menu “File Save” or “File Save as”).  

While in offloading process, a progress bar is displayed. The offload 
time is proportional to the amount of data being transferred (number 
of records in memory). 

After data transfer is completed, the 

Graph

 window is displayed.  

NOTE

: The logging is not stopped during the data offload process. 

The LogChart-II software includes a help section with detailed usage 
information. 

VISUALIZING THE DATA 

Three views are provided for data visualization. The records can be 
analyzed in the form of Chart (graph) or Table. A third window, called 

General  Information 

window  (

Fig.  15

),  shows  acquisition  details 

and logger configuration. 

A  region  of  interest  in  the  data  can  be  selected  (see  item 

Gr

aph 

Window

,  below).  The  Graph  Window  can  be  printed  and  the 

corresponding data exported as text files to other applications, such 
as spread sheets.  

The  same  windows  are  created  when  the  data  is  read  from  a 
previously saved file. 

After the windows are opened, one can save the data to a file (.lch), 
print  graphs,  or  export  the  data  from  the 

General  Information 

and 

Acquisitions Table

 windows

 

to a text file (.txt or .dat). 

Graph window 

The  Graph  is  a  convenient  tool  for  analysis.  As  one  moves  the 
mouse  in  the  chart  area,  the  time  and  the  value  of  the  records  of 
each  channel  are  shown  in  the  field  located  in  the  bottom  of  the 
window.  

Zooming in and out are implemented. They are available in the menu 
“View” or in the tool bar as icons. It is possible to select an area by 
clicking  and  dragging  the  mouse,  creating  this  way  a  zoom  region, 
starting at the upper left position of the region of interest.  

The  graphs  can  be  moved  vertically  by  moving  the  mouse  up  or 
down while keeping the mouse right button pressed. 

Fig. 14

 shows the Graph window. 

The Graph window is automatically opened after the offload process. 
The  two  other  windows  (Table  and  General  Information)  can  be 
opened accessing the menu or clicking on the corresponding button.

 

Fig. 14

 – Screenshot showing the graphic visualization of logged data

Содержание LogBox-DA

Страница 1: ...20mA 4 20mA 0 50mV or 0 5V user defined Note For the channel 2 depending on the input signal selected it may be necessary to configure an internal jumper DATA ACQUISITION LOGGING Data can be acquired by means of two different modes In channel 1 which counts pulses the number of input pulses counted in the logging time interval is stored in the logger memory at the end of the interval The maximum n...

Страница 2: ... the optical interface communication COM1 COM2 or USB CONFIGURING THE LOGGER Make sure the IR LINK3 wand is connected to the PC port selected The interface must be pointed towards the logger communication window see Fig 5 at a distance of about 15 cm Fig 5 Infrared interface position Click the button to start the communication between the logger and the software the Parameters Configuration window...

Страница 3: ... triggered Once activated the alarm LED indicator stays so even after the alarm triggering situation has ceased CHANNEL 2 ANALOG INPUT The options provided for Channel 2 configurations are Tag Defines a name up to 8 characters for identifying the variable to be measured Input The signal applied to the logger second input is defined here The following options are offered None disabled 4 20 mA 0 20 ...

Страница 4: ...he logger Note The case cover should not be opened unless battery replacement is required If this is the case the cover must be properly tightened back in its place such as to assure the IP67 68 protection Fig 9 IP67 external connectors Wire connections Channels cable right Connection Blue Channel 1 3 IN1 positive Black Channel 1 4 GND negative White Channel 2 1 IN2 positive Brown Channel 2 2 GND ...

Страница 5: ...logger during the process Data can be collected anytime and saved in files for future analysis menu File Save or File Save as While in offloading process a progress bar is displayed The offload time is proportional to the amount of data being transferred number of records in memory After data transfer is completed the Graph window is displayed NOTE The logging is not stopped during the data offloa...

Страница 6: ...rds the center of the device Worn batteries should not be recharged dismantled or incinerated After use batteries must be disposed according to local legal rules or returned to Novus After placing batteries back to the logger make sure the cover is firmly attached to the socket MOST FREQUENTLY PROBLEMS The LED is not flashing The LED flashing light is intentionally weak and it can be difficult to ...

Страница 7: ...mV 10 MΩ 0 10 Vdc 120 kΩ Memory capacity 64000 64k Interval between readings Minimum 1 second maximum 18 hours External battery switch time 0 to 10 s Power 3 6 V lithium battery 1 2 AA Estimated autonomy 200 days with one weekly download and 5 minutes measuring interval Battery life depends heavily on data retrieval frequency Working temperature From 40 C to 70 C Protection IP65 and IP67 models se...

Отзывы: