LogBox-DA Datalogger 

 

NOVUS AUTOMATION 

4/7

 

Note: Make sure that the compress  fitting is perfectly tightening the 
cables,  thus  assuring  proper  IP65  protection  (totally  dust-tight  and 
protected against water jets). 

 

 

Fig. 8

 – IP65 connections internal view 

CONNECTIONS IN THE IP67 MODEL 

In theIP67 version, a M8 connector is provided for signal input, as in 

Fig.  9

.  When  required,  a  second  M8  connector  can  be  installed  to 

output  the  signal  for  commanding  the  external  power  supply.  The 
cables are supplied with the logger. 

Note:  The  case  cover  should  not  be  opened  unless  battery 
replacement  is  required.  If  this  is  the  case,  the  cover  must  be 
properly tightened back in its place such as to assure the IP67 / 68 
protection.   

 

Fig. 9

 – IP67 external connectors

 

Wire connections 

Channels cable (right)

 

Connection 

Blue 

Channel 1 – (3 IN1) – positive (+) 

Black

 

Channel 1 – (4 GND) – negative (-) 

White

 

Channel 2 – (1 IN2) – positive (+) 

Brown

 

Channel 2 – (2 GND) – negative (-) 

 

Switch cable (left)

 

Connection 

Blue 

Transducer  (7 +VTRANS) 

Black

 

External Battery (5 -BAT) 

White

 

External Battery (6 +BAT) 

Brown

 

Not connected 

 

   

Fig. 10

 – Voltage pulse and dry contact channel 1 input 

 

Fig. 11

 – NPN and PNP open collector input for channel 1 

   

 

Fig. 12

 – Current and voltage inputs for channel 2 

Содержание LogBox-DA

Страница 1: ...20mA 4 20mA 0 50mV or 0 5V user defined Note For the channel 2 depending on the input signal selected it may be necessary to configure an internal jumper DATA ACQUISITION LOGGING Data can be acquired by means of two different modes In channel 1 which counts pulses the number of input pulses counted in the logging time interval is stored in the logger memory at the end of the interval The maximum n...

Страница 2: ... the optical interface communication COM1 COM2 or USB CONFIGURING THE LOGGER Make sure the IR LINK3 wand is connected to the PC port selected The interface must be pointed towards the logger communication window see Fig 5 at a distance of about 15 cm Fig 5 Infrared interface position Click the button to start the communication between the logger and the software the Parameters Configuration window...

Страница 3: ... triggered Once activated the alarm LED indicator stays so even after the alarm triggering situation has ceased CHANNEL 2 ANALOG INPUT The options provided for Channel 2 configurations are Tag Defines a name up to 8 characters for identifying the variable to be measured Input The signal applied to the logger second input is defined here The following options are offered None disabled 4 20 mA 0 20 ...

Страница 4: ...he logger Note The case cover should not be opened unless battery replacement is required If this is the case the cover must be properly tightened back in its place such as to assure the IP67 68 protection Fig 9 IP67 external connectors Wire connections Channels cable right Connection Blue Channel 1 3 IN1 positive Black Channel 1 4 GND negative White Channel 2 1 IN2 positive Brown Channel 2 2 GND ...

Страница 5: ...logger during the process Data can be collected anytime and saved in files for future analysis menu File Save or File Save as While in offloading process a progress bar is displayed The offload time is proportional to the amount of data being transferred number of records in memory After data transfer is completed the Graph window is displayed NOTE The logging is not stopped during the data offloa...

Страница 6: ...rds the center of the device Worn batteries should not be recharged dismantled or incinerated After use batteries must be disposed according to local legal rules or returned to Novus After placing batteries back to the logger make sure the cover is firmly attached to the socket MOST FREQUENTLY PROBLEMS The LED is not flashing The LED flashing light is intentionally weak and it can be difficult to ...

Страница 7: ...mV 10 MΩ 0 10 Vdc 120 kΩ Memory capacity 64000 64k Interval between readings Minimum 1 second maximum 18 hours External battery switch time 0 to 10 s Power 3 6 V lithium battery 1 2 AA Estimated autonomy 200 days with one weekly download and 5 minutes measuring interval Battery life depends heavily on data retrieval frequency Working temperature From 40 C to 70 C Protection IP65 and IP67 models se...

Отзывы: