CHEMOTRONIC III
55
Standard test method for cloud point
measurements
Apparatus:
1)
CHEMOTRONIC automatic cloud point tester
2)
Digital balance min. range 0..100 grams with 0.01
grams resolution
Preparation:
A 10% solution of the resin under test oil with the
high-quality grade mineral test oil must be prepared
as follow:
1)
Place the glass test tube with the magnet stirrer rod
into the test tube holder and place the holder on the
balance.
2)
Zero the balance.
3)
Weigh in 2.0 grams of resin.
4)
Zero the balance.
5)
Weigh in 18.0 grams of test oil.