9.3. The calibration menu
This menus is available within the Lab or PC mode only. In this menu the two global
reference points of the used turbidity range can be determined. Within the main
menu press the
cal
button to
enter
into the cal menu, the first reference liquid
(water or mineral distillate) is expected. For details, see "turbidity calibration".
9.1. The data menu
To simplify the report and the comparison of different measurement series a data
logger was implemented within the CHEMOTRONIC III. The data logger records the
relevant parameters of a successful turbidity determination in a special data sheet.
The following parameters will be saved:
1. The cloud point temperature
2. Time and date
3. Used program (Standard or U1..8)
4. A user definable comment
(only available via USB and the WinTurbi V2.0 software tool)
The data logger administrates 20 data sheets. After every successful turbidity
determination the measurement parameters are saved automatically and an index
counter for the next saving is incremented. If the counter value is equal 20 the
next value amounts 1. Attention, in this case the first data sheet will be over
written. The index of the last filled data sheet is displayed at left side of program
memory display by changing into data menu.