CHEMOTRONIC III
15
5.
Installation
The CHEMOTRONIC III is an automatic test instrument for table top use. It includes
all required mechanic and electronic components and requires no installation
experience.
When the CHEMOTRONIC III is put into operation the following two must be made:
1. The main cable for power supply. The standard operating voltage is
230 V ±
10 % / 50 – 60 Hz
. In case of
100 – 125 V mains voltage
, an optional
transformer is needed to increase the mains voltage to the standard operating
voltage of the instrument. Please note, always an
earth connection
is
available at the wall socket.
2. The Pt 100 temperature sensor with cable, protected by a spiralized metal
sheath, and a five pole screw type plug must be connected to the instrument.
The Pt 100 input socket is located at the rear side. Please note, there is a slot
in the input socket in which the nose of the plug has to fit. Place the plug on
the socket in the correct position first, before turning the knurled metal collar
of the plug clockwise in order to insert the plug into the socket.
Do not use a
lot of force and do not try to connect the plug in an incorrect position
.
This may destroy the fine gold-plated pins of the socket, so that the instrument
will need to be repaired not under guarantee.
To ensure a save operation of the device the user has to be sure to have a minimal
distance 25 of centimetres to any object at the rear side. When the connections
have been made, the equipment can be switched on by pressing the power push
button at the right side of front panel. If the self test proceeds without any failure
the led of the
start/stop
button is illuminating after two seconds.