CHEMOTRONIC III
25
9.
Principle of automatic controlled
measurement
For exact cloud point determinations it is very important to dissolve the hard resins
under reproducible circumstances. This means that in order to obtain reliable
results, it is always necessary to granted reproducible the following criteria:
- the rate of heating up the liquid
- the rate of cooling down the liquid
- the hold time at a high temperature level
- the stirring speed during the process
- the sensitivity of the photocell system
The following figure shows a typical measurement program:
Program and Parameter Description
T
t
Hold Temperature,
called Temp1
(adjustable)
Hold Temperature
– delta Temperature
Rotation Speed
Setpoint Speed1
(adjustable)
Temperature Profile
Setpoint Speed2
(adjustable)
Setpoint Speed3
(adjustable)
Speed Profile
Temperature of
Starting Speed3,
called Temp2
(adjustable)
heating up (hu)
T1, N1
Turbidity recognition activiation (level adjustable)
hold Temp (ho)
T1, N1
(Time adjustable)
cooling down (cd1)
T2, N2
cooling down (cd2)
T3, N2
cooling down (cd3)
T3, N3
(T3 adjustable)