CHEMOTRONIC III
33
Print an error message:
If one or more of the following errors are detected the current measurement will be
stopped and the restart will be disabled. In addition the alarm led is blinking and an
error code is shown within the program display. Simply press the
enter
button to
get a detailed failure report by the line printer.
The coded errors showed below will be detected:
1. Not plausible temperature gradient
E 1
2. Stirring motor is out of order
E 2
3. At least one cooling motor is out of order
E 4
4. The temperature within the heating channels is to high
E 16, E32
5. The temperature within the cooling box is to high
E 64