FOCUSING OPTICS
PAGE
|
35
Table 3-1a Dew point temperatures in Celsius or Centigrade.
Dew Point Table °C
Dew point
The dew point table provides dew point temperatures for a range of air temperature and relative
humidity values. Remember that the laser’s coolant temperature must be set above the dew point
temperatures given in the chart but should not exceed 22 °C (72 °F).
To use the table, look down the Air Temp column and locate an air temperature in Fahrenheit or Celsius
(°C values are shown in parentheses) that corresponds to the air temperature in the area where your
laser is operating. Follow this row across until you reach a column matching the relative humidity in your
location. The value at the intersection of the Air Temp and Relative Humidity columns is the Dew Point
temperature in °F (or °C).
The chiller’s temperature set-point must be set above the dew point temperature. For example, if the air
temperature is 85 °F (29 °C) and the relative humidity is 60%, then the dew point temperature is 70 °F
(21 °C). Adjust the chiller’s temperature set-point to 72 °F (22 °C) to prevent condensation from forming
inside the laser or RF power supply.
Содержание SYNRAD Pulstar p100
Страница 1: ...ENGINEERED BY SYNRAD p100 150 Lasers User Manual...
Страница 17: ...NOMENCLATURE CONTINUED PAGE 12 Nomenclature Continued Figure 1 2 Anatomy of a model number...
Страница 23: ...ADDITIONAL LASER SAFETY INFORMATION PAGE 18 p100 label locations...
Страница 24: ...ADDITIONAL LASER SAFETY INFORMATION PAGE 19 p150 label locations...
Страница 30: ...ELECTROMAGNETIC INTERFERENCE STANDARDS PAGE 25 Figure 2 1 p100 Declaration Document...
Страница 31: ...ELECTROMAGNETIC INTERFERENCE STANDARDS PAGE 26 Figure 2 2 p150 Declaration Document...
Страница 39: ...FOCUSING OPTICS PAGE 34 Table 3 1a Dew point temperatures in Fahrenheit Dew Point Table F...
Страница 46: ...OPERATION MODES PAGE 41 Figure 3 6 p150 pulse profile 37 5 duty cycle at 10 kHz...
Страница 49: ...USER I O CONNECTION SUMMARY PAGE 44 Figure 3 9 User I O connector pinouts...
Страница 68: ...REMOTE INTERLOCK FUNCTIONS PAGE 63 Figure 3 21 p100 150 outline mounting drawings pg 1 of 2...
Страница 70: ...REMOTE INTERLOCK FUNCTIONS PAGE 65 Figure 3 23 p100 150 packaging instructions...
Страница 71: ...REMOTE INTERLOCK FUNCTIONS PAGE 66 Figure 3 24 p150 outline and mounting...
Страница 78: ...OPERATIONAL FLOW CHART PAGE 73 Operational flow chart Figure 4 1 p100 150 operational flowchart...
Страница 79: ...STATUS LEDS PAGE 74 Status LEDs Table 4 2 p100 p150 Input output LED Status Signals...
Страница 88: ...This page is left blank intentionally...