38
Chapter 7
Technical specifications
The lower and upper limits of the linear operating range
Tab. 2 The lower and upper limits of the linear operating range
A-weighting
C-weighting
Sound level (dB)
Frequency (Hz) 31.5
1 k
4 k
8 k
12.5 K
Upper
97.0
137.0
136.0
133.0
130.0
Start
94.0
94.0
94.0
94.0
94.0
Lower
30.0
30.0
30.0
30.0
30.0
Sound level (dB)
Frequency (Hz) 31.5
1 k
4 k
8 k
12.5 K
Upper
134.0
137.0
136.0
133.0
130.0
Start
94.0
94.0
94.0
94.0
94.0
Lower
36.0
36.0
36.0
36.0
36.0