20
Chapter 7
Technical specifications
Inherent noise level
Wide range: A weighting: 21 dB or less
C weighting: 29 dB or less
Peak range: A weighting: 54 dB or less
C weighting: 54 dB or less
Measurement frequency range
10 Hz to 20 kHz
Reference frequency
1 kHz
Reference sound pressure level
94 dB
Frequency weighting characteristics
A and C
Time weighting characteristics
F (fast) and S (slow)
Level range
Wide range: 30 to 130 dB
Peak range*: 60 to 130 dB
* Peak range is used when measuring peak sound level.
RMS detecting circuit
Digital processing
Processing
Digital
Sampling interval: 25 µs
(L
p
, L
eq
, L
max
, L
E
, L
Cpeak
)
Calibration acc IEC, JIS:
Acoustic calibration using Nor1255 or similar
Calibration frequency:
1 kHz
Calibration sound pressure level: 114 dB
Electronic verification using an internal electronic signal