28
Chapter 7
Technical specifications
6.2.2
(Note)
Measurement when static pres-
sure is < 85 kPa
Calibration and measurement
performed in this environment
using Sound Calibrator Nor1255
6.5.2 Degradation of functions
by electrostatic discharge
9.2.7 b) Measurement value
6.5.2
Degradation of functions by
electrostatic discharge
9.2.7 b)
Measurement value affected
temporarily by electrostatic
discharge
6.6.1
Operating mode with least
immunity to AC power frequency
fields and RF fields
9.3 o)
Fig. 5 Operation mode: normal
operation
Connection pattern: PC adapter
cable Nor4596 (with ferrite
cores)
6.6.4
(Note)
Field strength for conforming
(in case > 10 V/m)
9.3 m)
N/A
7
Provisions for use with auxiliary devices
7.1
Correction for use of microphone
cable
9.2.6 b)
N/A
7.2
Effect of optional accessories
(windscreen)
9.2.6 a)
Fig. 4 Influence of Nor4617
on acoustic performance of
Nor103
7.3
Statement of conformance with
optional accessories (windscreen)
Compliant with IEC 61672-1,
with Windscreen Nor4617
mounted
7.4
Operation of 1/1 - 1/3 octave band
filters
9.2.6 c)
N/A
7.5
Details about connection & effects
of auxiliary devices
9.2.6 d)
Attachments
9
Instruction manual
9.2.1
General
9.2.1 a)
Description of type, classifi cation
(X, Y, Z) and class
Group X, Class 1
9.2.1 b)
Overall confi guration,
Normal operation confi guration
(including windscreen)
5.1.4
5.1.7
Refer to 5.1.4
Refer to 5.1.7
9.2.1 c)
Models of microphones
5.1.6
Refer to 5.1.6
9.2.1 d)
Required microphone cable to
conform
N/A
9.2.1 e)
Characteristics & operation each
channel
N/A
9.2.2
Design features