88
8.1 A
CONNECTOR MEASUREMENT PROCEDURE
STEP 1
•
Obtain the following:
•
Computer and Monitor powered up with NCC software running.
•
CC 6000 Microscope
•
Mount
•
PC - Specific to the ferrule size of the connectors being measured.
OR
•
APC - Specific to the connector type (i.e. FC, ST, or LC)
•
Mount screws
•
Alcohol & Lint free wipes or the supplied connector cleaner
•
Connectors to be analyzed.
* This procedure can also be used for Customer specific mounts.
STEP 2
•
Calibrate the mount:
•
Use Reference Tool (or Leveling Connector) Calibration Procedure for PC mounts
•
Use Reference Tool Calibration Procedure for APC mounts
•
Set rear panel Mode switch according to Ferrule Material.
•
Mode 1 for Ceramic ferrule.
•
Mode 2 for Metal ferrule.