76
7.3
CALIBRATION TAB
The Calibration Tab contains
all needed calibration features to
assure the microscope
measures accurately. Some
features are hidden and can only
be used by administrator level
users. All Calibrations should be
done the microscope’s Mode
switch (on its back panel) set to
position 1
Offset Calibration Procedure:
Offset Calibration procedure should be conducted after every mount change. There
are two types available -- “Reference Tools” and “Leveling Connector”. Which of these
types to perform depends on the calibration artifact owned as well as the desired
accuracy of calibration needed.
A Reference Tool, with known endface geometry parameters, is supplied with the
microscope and assures that the interferometer measures accurately. The known X and
Y angular offset angles are compared to an actual measurement taken during the
Calibration procedure. The software calculates the calibration factors so the known
values are equal to the actual measurement values.
There are PC and APC reference tools. The PC reference tool is used in unison with
a PC Locking V Groove Mount. The APC reference tool is used in unison with an APC
Locking V Groove Mount.
•
Reference Tools tab
:
In this Reference Tools tab, a preset Reference Tool ID name can be selected from
the list of choices, with each ID’s values displayed. These preset Reference Tool
ID choices can be constructed and saved for use in this tab using the
“Tools\Reference Tools Add_Edit” menu choice. After selecting a Reference Tool
choice, insert that tool into the CC6000 mount (with its arm leaning against the left
protruding bar), adjust the front panel focus knob until best contrast fringes can be
seen on the Live View, and click the Calibrate button to proceed with calculation
CALIBRATION TAB
Содержание Connect-Chek CC6000
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 13: ......
Страница 14: ......
Страница 16: ......
Страница 28: ...C H A P T E R 2 28 SYSTEM PREPARATION...
Страница 33: ...33 2 3 SOFTWARE INSTALLATION 5 Click the Next button for the following three windows that appear...
Страница 40: ...C H A P T E R 3 40 NCC PROGRAM...
Страница 42: ...42 3 1...
Страница 43: ...C H A P T E R 4 43 MENU BAR FUNCTIONS...
Страница 45: ...45 4 1...
Страница 55: ...55 4 4...
Страница 56: ...56 C H A P T E R 5 TOOLBAR FUNCTIONS...
Страница 64: ...64 C H A P T E R 6 IMAGE TABS...
Страница 67: ...C H A P T E R 7 67 CONTROL TABS...
Страница 75: ...75 7 2 SCAN OPTIONS TAB...
Страница 82: ...82 7 4 REFERENCE TOOL CALIBRATION...
Страница 92: ...C H A P T E R 9 92 APPENDIX...
Страница 104: ...104 9 3 SHIPPING INSTRUCTIONS...