24
1.4
Background:
Because of their precise nature, interferometric measurements are
sensitive to vibration. Vibration can occur from rotating machinery, nearby traffic,
acoustic noise and a variety of other sources. Touching the instrument or the surfaces on
which it sits during measurements can also affect the measurements.
Instrument placement considerations:
The instrument should be placed on a solid
support away from obvious sources of vibration (e.g. rotating machinery). Ideally, the
instrument should rest on a separate surface so that vibrations from using the computer
keyboard and vibrations from other activities are not transferred through the surface to
the instrument.
Vibration Isolators:
If the vibration is present even after the instrument is located on
its own surface away from vibration sources, then additional vibration isolation needs to
be employed. A low cost solution for vibration isolation is to place the instrument on a
piece of vibration isolation material such as Sorbothane. More sophisticated vibration
isolation tables are available from manufacturers such as Newport, Herzan, and
Technical Manufacturing Corporation (TMC). Before purchasing equipment, a technical
representative of the isolation equipment supplier should visit the site where the
instrument is installed to recommend the appropriate equipment to isolate the instrument
from the vibration sources present.
VIBRATION ISSUES
Содержание Connect-Chek CC6000
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 13: ......
Страница 14: ......
Страница 16: ......
Страница 28: ...C H A P T E R 2 28 SYSTEM PREPARATION...
Страница 33: ...33 2 3 SOFTWARE INSTALLATION 5 Click the Next button for the following three windows that appear...
Страница 40: ...C H A P T E R 3 40 NCC PROGRAM...
Страница 42: ...42 3 1...
Страница 43: ...C H A P T E R 4 43 MENU BAR FUNCTIONS...
Страница 45: ...45 4 1...
Страница 55: ...55 4 4...
Страница 56: ...56 C H A P T E R 5 TOOLBAR FUNCTIONS...
Страница 64: ...64 C H A P T E R 6 IMAGE TABS...
Страница 67: ...C H A P T E R 7 67 CONTROL TABS...
Страница 75: ...75 7 2 SCAN OPTIONS TAB...
Страница 82: ...82 7 4 REFERENCE TOOL CALIBRATION...
Страница 92: ...C H A P T E R 9 92 APPENDIX...
Страница 104: ...104 9 3 SHIPPING INSTRUCTIONS...