Robot Initial Setup (Step 7 Tab): Testing the System Setup and Calibration (Non-Laser Systems
Only)
#
Click
Step
Reference Image
1
• Click the STEP7 tab.
2
• Click NEEDLE Z DETECT to test the setup.
• Click YES/OK when prompted for
confirmations.
NOTES:
- When the system performs a Needle Z
Detect, it automatically performs a Needle
XY Adjust directly after performing the
Needle Z Detect.
- Refer to “How the System Responds to
Needle Z Detect or Needle XY Adjust”
below for a detailed description of the
system response to a Needle Z Detect
selection.
3
>
• Click SAVE.
• Click FINISH.
The system is now properly set up and
calibrated. Refer to “Programming” on
page 63 to create programs.
Setting Up the System Using the Robot Initial Setup Wizard (continued)
How the System Responds to Needle Z Detect or Needle XY Adjust
When you click NEEDLE Z DETECT, the system performs the following actions:
• Moves the dispensing tip over the tip detector sensor and lowers it until it touches the sensor.
• Measures and compares the difference between the last measurement and the current measurement.
• Requests confirmation for any change in the tip-to-workpiece offset (Z clearance).
• Realigns all points in the currently open program to the new tip-to-workpiece offset (Z clearance).
• Automatically performs a Needle XY Adjust sequence (shown below).
When you click NEEDLE XY ADJUST, the system performs the following actions:
• Moves the dispensing tip to a preset location on the workpiece.
• Dispenses a dot of fluid.
• Moves the camera over the deposited dot of fluid.
• Compares the alignment of the dot with the mark image saved in the Mark Library.
• Requests confirmation for any change in the laser-to-tip (if applicable) or camera-to-tip offset (XY offsets).
• Realigns all points in the currently open program to the new XY offsets.