![Nonin CO-Pilot H500 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/nonin/co-pilot-h500/co-pilot-h500_operators-manual_1700084058.webp)
52
Technical Information
Equipment Response Time
If the SpO
2
signal from the sensor is inadequate, the last measured values
freeze for 10 seconds and are then replaced with dashes.
*Wireless communications can be delayed when in environments with high
radio interference.
Example – SpO
2
Exponential Averaging
SpO
2
decreases 1.0% every 2 seconds (5% over 10 seconds)
Pulse Rate = 75 BPM
Specific to this example:
• The response of the SpO
2
average is 6 seconds.
SpO
2
Values
Response
Latency
Fast Averaged SpO
2
3 second or faster exponential time constant
2 beats
Pulse Rate Values
Response
Latency
Fast Averaged
Pulse Rate
3 second or faster exponential time constant
2 beats
Equipment Delays
Delay
Display Update Delay
1.5 – 2.5 seconds*
Содержание CO-Pilot H500
Страница 1: ...Operator s Manual CO Pilot Model H500 Wireless Handheld Multi Parameter System English...
Страница 18: ...12 Display Screens and Symbols Figure 2 CO Met Measurement Screen CO Met Disabled 1 3 7 2 9 10 13 12 11 19...
Страница 19: ...13 Display Screens and Symbols Pulse Oximetry Figure 3 SpO2 Measurement Screen 2 3 7 10 11 13 12 1 9...
Страница 20: ...14 Display Screens and Symbols Cerebral and Tissue Oximetry Figure 4 rSO2 Measurement Screen 1 5 8 11 12...
Страница 37: ...31 System Operation Figure 17 Data Transfer Successful...
Страница 66: ...60 Technical Information THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 67: ...61 Technical Information THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 68: ......