44
Troubleshooting
Error Codes
Certain operation conditions may cause error codes to pop-up on screen. These
error codes are detailed below:
Error Code
Possible Cause
Possible Solution
COM-01
Wireless connection setup
failed. Connection may have
taken too long to open.
Attempt connection again.
If problem persists, contact
Nonin.
COM-02
Wireless connection
rejected, likely due to pairing
mismatch.
Contact Nonin.
COM-03
Wireless connection failed
security authentication.
Attempt connection again. If
problem persists, contact Nonin.
OXI-01
SP-BLE is not sending
critical data.
Attempt connection again.
Power the device off and then on.
If problem persists, contact
Nonin.
OXI-02
SP-BLE internal error.
Power the device off and then on.
If problem persists, contact
Nonin.
OXI-03
SP-BLE gas gauge failure.
Power the device off and then on.
If problem persists, contact
Nonin.
MEM-01
Application configuration
failure.
Contact Nonin.
MEM-02
Failure during file transfer.
OTG cable may have been
bumped during file transfer.
Attempt connection again.
If problem persists, contact
Nonin.
MEM-03
Failure during file deletion.
Attempt connection again.
If problem persists, contact
Nonin.
MEM-04
Failure creating study file.
Attempt connection again.
If problem persists, contact
Nonin.
Содержание CO-Pilot H500
Страница 1: ...Operator s Manual CO Pilot Model H500 Wireless Handheld Multi Parameter System English...
Страница 18: ...12 Display Screens and Symbols Figure 2 CO Met Measurement Screen CO Met Disabled 1 3 7 2 9 10 13 12 11 19...
Страница 19: ...13 Display Screens and Symbols Pulse Oximetry Figure 3 SpO2 Measurement Screen 2 3 7 10 11 13 12 1 9...
Страница 20: ...14 Display Screens and Symbols Cerebral and Tissue Oximetry Figure 4 rSO2 Measurement Screen 1 5 8 11 12...
Страница 37: ...31 System Operation Figure 17 Data Transfer Successful...
Страница 66: ...60 Technical Information THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 67: ...61 Technical Information THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 68: ......