Table 7. Electromagnetic Immunity
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment—Guidance
This device is intended for use in the electromagnetic environment specified below. The
customer and/or user of this device should ensure that it is used in such an environment.
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
relative humidity should be at
least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
N/A
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
N/A
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
±5% U
T
(>95% dip
in U
T
)
for 0.5 cycle
±40% U
T
(60% dip
in U
T
)
for 5 cycles
±70% U
T
(30% dip
in U
T
)
for 25 cycles
<5% U
T
(>95% dip
in U
T
)
for 5 seconds
N/A
Mains power quality should be
that of a typical commercial or
hospital environment.
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE:
U
T
is the AC mains voltage before application of the test level.
Содержание CO-Pilot H500
Страница 1: ...Operator s Manual CO Pilot Model H500 Wireless Handheld Multi Parameter System English...
Страница 18: ...12 Display Screens and Symbols Figure 2 CO Met Measurement Screen CO Met Disabled 1 3 7 2 9 10 13 12 11 19...
Страница 19: ...13 Display Screens and Symbols Pulse Oximetry Figure 3 SpO2 Measurement Screen 2 3 7 10 11 13 12 1 9...
Страница 20: ...14 Display Screens and Symbols Cerebral and Tissue Oximetry Figure 4 rSO2 Measurement Screen 1 5 8 11 12...
Страница 37: ...31 System Operation Figure 17 Data Transfer Successful...
Страница 66: ...60 Technical Information THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 67: ...61 Technical Information THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 68: ......