25
System Operation
5.
Once display has connected to signal processor, measurement screen
will display. The system is now ready to take a measurement.
NOTE:
If the sensor does not turn on, ensure sensor is plugged into the
signal processor. If the sensor still does not turn on, press and hold down the
On/Off, Sleep
button on the signal processor.
Taking a Measurement
1.
Prepare the system for measurement by following the steps in the
previous section, “Start-up Sequence”.
2.
Place the sensor on the patient according to sensor instructions.
3.
Take measurement.
4.
When operation is complete, briefly press the
On/Off, Sleep
button on
display to end session and turn off the device. Signal processor will turn
off automatically.
During patient measurements the system may reach temperatures of up to 43
degrees Celsius, which is safe for contact with healthy adult skin (no time
restriction).
Determining Pulse Signal Quality
The pulse signal quality can be determined by the color of the pulse waveform
displayed on the display. This pulse waveform changes colors to alert you to
changes in pulse quality that may affect the readings:
•
Green
Pulse Waveform indicates a good pulse signal.
•
Yellow
Pulse waveform indicates a marginal pulse signal.
•
Red
Pulse Waveform indicates an inadequate pulse signal.
Figure 11. Pulse Signal Quality
Green = Good
Yellow = Marginal
Red = Insufficient
Содержание CO-Pilot H500
Страница 1: ...Operator s Manual CO Pilot Model H500 Wireless Handheld Multi Parameter System English...
Страница 18: ...12 Display Screens and Symbols Figure 2 CO Met Measurement Screen CO Met Disabled 1 3 7 2 9 10 13 12 11 19...
Страница 19: ...13 Display Screens and Symbols Pulse Oximetry Figure 3 SpO2 Measurement Screen 2 3 7 10 11 13 12 1 9...
Страница 20: ...14 Display Screens and Symbols Cerebral and Tissue Oximetry Figure 4 rSO2 Measurement Screen 1 5 8 11 12...
Страница 37: ...31 System Operation Figure 17 Data Transfer Successful...
Страница 66: ...60 Technical Information THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 67: ...61 Technical Information THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 68: ......