NOISE COM INC
After a brief delay, the MAIN INFORMATION DISPLAY will
appear, similar to the display shown below:
MANUAL MODE, on the top line, indicates that the PNG is
ready to accept commands from the KEYPAD and has received
no commands from a Remote Interface (GPIB, RS-232, RS-
485). If the most recent command to the PNG were received
from a Remote Interface, rather than the Keypad, the top
line would indicate 'REMOTE MODE' to advise the user that
the PNG has most recently been under Remote Control, and
that a Keypad action (if not disabled by a remotely issued
Local Lockout Command) may interrupt a remotely controlled
operation.
If the basic unit was ordered without options then the
display would indicate the noise attenuation of 127 dB
(unless a >2 GHz unit then 79 dB is the maximum attenuation)
and the noise filter at position 1.
The internal Noise Attenuator, 'NA', and Signal Attenuator
(for an externally applied signal), 'SA' are each set to
the maximum attenuation permitted by the PNG Type. The
internal Noise Filter, 'NF', and Signal Filter, 'SF', are
set to position #1. Position #1 is a through path full
(bandwidth). The Step Size, 'ST', is set to the minimum
increment permitted by the PNG Type. If the PNG has
been internally configured to support a Power Meter, the
Noise Power delivered to the load will be displayed in this
position.
Document M016301
Pg 11
Rev. 2 2/26/03
MANUAL MODE
NA=127.9dB NF=1
SA=127.9dB SF=1
ST=000.1 P=000.0dB
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com