The following sequence of instructions enters the same sequence
of program steps described in the Keypad Programming example
above. Note that specific 2- character codes are used to
represent Program Functions, and that all such Program Entry
Commands must have the same opening format 'Pn:mm '
P0:00 NS002.0 cr
P0:08 NAI cr
P0:01 NF0002 cr
P0:09 TD 5 cr
P0:02 NA 0 cr
P0:10 NAI cr
P0:03 TD 5 cr
P0:11 TD 5 cr
P0:04 NAI cr
P0:12 NAI cr
P0:05 TD 5 cr
P0:13 TD 5 cr
P0:06 NAI cr
P0:14 GO 0002 cr
P0:07 TD 5 cr
P0:15 EX
Where cr is either a cr character for the RS-232 option
or for GPIB interface operation it is a LF or EOI hard-
ware or both.
Valid 2-character Function Codes for PNG-III Programs
are:
NA := Noise Attenuator
XA := eXternal signal Attenuator
NF := Noise Filter
XF := eXternal signal Filter
NS := Noise Step
XS := eXternal signal Step
TD := Time Delay
GO := GOTO
EX := EXIT
Note that it is possible to enter a Signal Step size that is
different than the Noise Step size, when programming via a
Remote Port. It is important to note that a Noise Step Command,
whether entered from the Keypad or a Remote Port, will set both
the Noise Step and Signal Step values. An 'XS' Command, entered
via a Remote Port, will set only the Signal Step value.
Therefore, to enable differing values for the two Step sizes,
the Signal Step value must be set AFTER setting the Noise Step
value.