NOISE COM, INC.
As digits are pressed at the keypad, they enter the parameter
display from the right and displace previously entered digits to the
left. The parameter display may be reset to zero ('0000' or '0') by
pressing the 'CLEAR' key.
Invalid entries will not be accepted. The PNG will automatically
limit the Attenuation Value from '000.0' to the maximum value
permitted by the PNGI Type (typically 127 or'127.9' or '079.9').
Valid Filter settings range from '1' to '8', inclusive.
Filter setting 1 represents full bandwidth ie a through
path. Filter setting 2 terminates the noise into a load internally,
ie this is a standby state. These are the only valid settings unless
the unit has been configured with the filter option. In this case
Filter settings 3-8 represent either internal filters ordered by the
customer or back panel connections enabling the customer to connect
his own external filters. In this case the first filter is
actualized with filter setting 3, the second filter is setting 4,
etc.
INCREMENTING / DECREMENTING ATTENUATORS
The Noise and Signal Attenuators may be incremented or decremented
by the programmed STEP size using only two key strokes. To use this
feature, first set the Step Size to the appropriate value.
Pressing the 'STEP' key will cause the following Information
Display to appear:
The numeric and 'ENTER' keys are used, in the same manner as for
setting Attenuators, to set the Step Size. The selected value will
be added to or subtracted from the selected Attenuator when the INC
and DEC Function Keys are used first in a sequence.
To execute this function, press either the 'INC' or 'DEC' key, as
appropriate, followed by the 'NOISE ATTEN' or 'SIGNAL ATTEN'
key. The new Attenuator setting will be reflected in the MAIN
INFORMATION DISPLAY, and a Cursor Bar will call attention to the
most recent change.
Document
M016301
Pg 14
Rev. 2 2/26/03
KEYBOARD SETUP
ATTEN STEP =000.1dB
NEW SETTING=000.0dB