WiMo / WiMo Plus Multiparameters Sondes
9
Operation
2.1.2
Install/Remove sensors
Unscrew the sensor locking sleeve in the anti-clockwise direction by hand or with the key tool
provided. Pull the sensor to extract it. Place it on a clean surface and store it taking care not to
damage the O-rings.
Before repositioning the sensor check that the location in the sonde has no damage and no
scratches. Grease the two sensor O-rings with the supplied grease and engage it in the location by
making the coded pin coincide with the host hole in the sonde. Take care to the sensor's centering in
its location. Move the sensor slightly from right to left to check that the pin is in front of its
receptacle before starting to screw. Be careful not to damage the threads of the sonde. Any
deterioration of the threads is not covered by the warranty.