WiMo / WiMo Plus Multiparameters Sondes
17
WEB Interface
3.1.2
Storage
There are two file formats: CSV or Topkapi's format TXT2.
It is possible to cut the measurement files to a size adjustable by the end-user. The goal is to avoid
losing data for long time deployments on a low battery shutdown for example. In operation with the
WiMo Modem, this option is automatically activated and the measurement files cut to a size of 256K.
Data files can be compressed to ZIP to save memory space.
3.1.3
Network and communication
Wi-Fi network settings can be changed, including the narrow bands used if they are cluttered.
To be able to communicate in Modbus with the sonde you must enable the Modbus function in the
interface. It will only work if an external power supply is detected. It is possible to select the RS232 or
RS485 support, as well as the address of the sonde (128 by default).
Wi-Fi network settings
Modbus RTU communication setting
Information on memory
Selecting the file format to use
File Settings in CSV mode
Option to cut out measurement files
Option to ZIP measurement files
Turning live view recording on/off